ID and susceptibility testing results for BCs (each species in six aerobic and six anaerobic bottles) artificially spiked with nine reference strains (phase II)

Species % Correct identification Results of susceptibility testing (%)a
Very major errors Major errors Minor errors
E. coli 100 0 0 0
E. coli (ESBLb positive) 100 0 0.8 2.5
P. aeruginosac 100 0 0 1.7
K. pneumoniae 100 0 0 2.9
K. oxytoca 100 0 0 5.0
S. rubidea 67 0 0 0.8
E. aerogenes 100 0 0 0
A. baumanniic 67 0 0 3.8
P. mirabilis 100 0 0.4 2.1
    Total 93.8 0 0.2 2.2