Table 1.
Mean Normalized Relative differences of GOx Systems 1–5 and GDH System 6 at pO2 Values <45 and ≥150 mmHg (20 Samples Each, 5 Measurements per Sample)a
| Oxygen sensitivity | GOx systems | GDH system | ||||
| Labeledb | Not labeledc | |||||
| System 1 | System 2 | System 3 | System 4 | System 5 | System 6 | |
| Relative differences pO2 <45 mmHg | ||||||
| Mean | 11.2% | 22.6% | 19.1% | 6.1% | 18.0% | -1.0% |
| Standard deviation | 3.8% | 6.2% | 8.7% | 2.3% | 6.4% | 1.7% |
| Minimum | 4.0% | 10.6% | 4.7% | 3.0% | 7.4% | -0.2% |
| Maximum | 19.2% | 32.1% | 37.9% | 13.0% | 29.0% | -4.5% |
| padjusted | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | 0.1505 |
| Relative differences pO2 ≥150 mmHg | ||||||
| Mean | -14.2% | -14.9% | -12.6% | -7.9% | -12.8% | -0.4% |
| Standard deviation | 3.8% | 3.8% | 3.0% | 2.0% | 4.8% | 1.6% |
| Minimum | -7.7% | -8.0% | -6.8% | -5.1% | -6.5% | 0.6% |
| Maximum | -24.3% | -23.2% | -17.3% | -14.4% | -28.9% | -3.4% |
| padjusted | <0.0001 | <0.0001 | <0.0001 | <0.0001 | <0.0001 | 1.0 |
The p values adjusted according to Bonferroni show whether normalized relative mean differences were significantly different from zero.
Sensitivity to increased blood pO2 was labeled in the package inserts.
Potential influences of pO2 on measurement results were not labeled in the user manual/package inserts.