Number of Device Failures (No Alarm and Wrong Reading Outside ±20%) as Documented for the Different Blood Volumes with the Different Blood Glucose Monitoring Systemsa

Device Sample volume
0.20 µl 0.32 µl 1.2 µl All
ACCU-CHEK Aviva 1/24 (4%) 0/24 (0%) 0/24 (0%) 1/72 (1%)
ACCU-CHEK Aviva Nano 0/24 (0%) 0/24 (0%) 8/24 (33%) 8/72 (11%)
ACCU-CHEK Compact+ 0/24 (0%) 0/24 (0%) 0/24 (0%) 0/72 (0%)
ACCU-CHEK Mobile 0/24 (0%) 0/24 (0%) 0/24 (0%) 0/72 (0%)
AlphaCheck+ 0/24 (0%) 0/24 (0%) 8/24 (33%) 8/72 (11%)
BGStar 0/24 (0%) 0/24 (0%) 0/24 (0%) 0/72 (0%)
iBGStar 0/24 (0%) 0/24 (0%) 0/24 (0%) 0/72 (0%)
Breeze 2 0/24 (0%) 0/24 (0%) 2/24 (8%) 2/72 (3%)
Contour 0/24 (0%) 1/24 (4%) 0/24 (0%) 1/72 (1%)
Contour USB 0/24 (0%) 0/24 (0%) 2/24 (8%) 2/72 (3%)
Elite 0/24 (0%) 0/24 (0%) 17/24 (71%) 17/72 (24%)
FineTouch 0/24 (0%) 0/24 (0%) 5/24 (21%) 5/72 (7%)
FreeStyle Freedom 1/24 (4%) 0/24 (0%) 0/24 (0%) 1/72 (1%)
FreeStyle Freedom Lite 0/24 (4%) 1/24 (4%) 0/24 (0%) 1/72 (1%)
GlucoCard+ 0/24 (0%) 4/24 (17%) 0/24 (0%) 4/72 (6%)
GlucoCard Meter 5/24 (21%) 17/24 (71%) 18/24 (75%) 40/72 (56%)
GlucoMen GM 0/24 (0%) 3/24 (13%) 0/24 (0%) 3/72 (4%)
GlucoMen LX 0/24 (0%) 10/24 (42%) 9/24 (38%) 19/72 (26%)
GlucoTalk 0/24 (0%) 0/24 (0%) 1/24 (4%) 1/72 (1%)
MediTouch 7/24 (29%) 3/24 (13%) 20/24 (83%) 30/72 (42%)
OmniTest 0/24 (0%) 1/24 (4%) 11/24 (46%) 12/72 (17%)
OneTouch Ultra Easy 0/24 (0%) 0/24 (0%) 6/24 (25%) 6/72 (8%)
OneTouch Ultra 2 0/24 (0%) 0/24 (0%) 3/24 (13%) 3/72 (4%)
OneTouch Ultra Smart 0/24 (0%) 0/24 (0%) 1/24 (4%) 1/72 (4%)
OneTouch Verio 0/24 (0%) 0/24 (0%) 0/24 (0%) 0/72 (0%)
OneTouch Vita 0/24 (0%) 1/24 (4%) 10/24 (42%) 11/72 (15%)
Precision Xceed 0/24 (0%) 0/24 (0%) 17/24 (0%) 17/72 (24%)
Precision Xtra 0/24 (0%) 2/24 (8%) 3/24 (13%) 5/72 (7%)
Wellion Calla Dialog 0/24 (0%) 0/24 (0%) 1/24 (4%) 1/72 (1%)
Wellion Calla Premium 0/24 (0%) 0/24 (0%) 3/24 (13%) 3/72 (4%)
Ypsomed Pura 0/24 (0%) 7/24 (29%) 20/24 (83%) 27/72 (38%)
a

Shadowed fields indicate the volumes with reported device failures.