Table 2.

Type of Device Response (Initiation of a Correct Reading or Low-Volume Alarm) in Those Devices with 100% Correct Response to the Artificially Provided Low Sample Volumes

Device Sample volume
0.20 µl 0.32 µl 1.2 µl All
Number of readings 24 24 24 72
BGStar
Low-volume alarm 8.0% 12.5% 0% 6.8%
Correct reading 92.0% 87.5% 100% 93.2%
iBGStar
Low-volume alarm 79.0% 62.5% 0% 47.2%
Correct reading 21.0% 37.5% 100% 52.8%
ACCU-CHEK Compact+
Low-volume alarm 100% 100% 66.5% 88.8%
Correct reading 0% 0% 33.5% 11.2%
ACCU-CHEK Mobile
Low-volume alarm 41.5% 29.0% 16.5% 29.0%
Correct reading 48.5% 71.0% 83.5% 71.0%
OneTouch Verio
Low-volume alarm 21.0% 4.0% 0% 8.3%
Correct reading 79.0% 96.0% 100% 91.7%