Table 3.
Summary of Analytical Performance Evaluation of the Test Strip (ISO 15197:2003 Section 7 and ISO 15197:2013 Section 6.3)
| Meter system | Reference method | Readings within specificed error limits (N = 600) | R2 | Parkes consensus error grid analysis | |
| ISO 15197:2003 criteriaa | R2 ISO 15197:2013 criteriab | ||||
| CXT/CNEZ16,33 | YSI | 100% | 100% | 0.9964 | 100% in zone A |
| CNL13,33 | YSI | 100% | 100% | 0.9961 | 100% in zone Ac |
| CNUSB14,33 | YSI | 100% | 100% | 0.9957 | 100% in zone A |
| CN15,33 | YSI | 100% | 99.8% | 0.9959 | 100% in zone A |
Within ISO 15197:2003 accuracy criteria (i.e., ≥95% of results shall fall within ±15 mg/dl or ±20% for samples with glucose concentrations < 75 and ≥75 mg/dl, respectively).
Within ISO 15197:2013 accuracy criteria (i.e., ≥95% of results shall fall within ±15 mg/dl or ±15% for samples with glucose concentrations <100 and ≥100 mg/dl, respectively).
Data on file.