Table 3.

Summary of Analytical Performance Evaluation of the Test Strip (ISO 15197:2003 Section 7 and ISO 15197:2013 Section 6.3)

Meter system Reference method Readings within specificed error limits (N = 600) R2 Parkes consensus error grid analysis
ISO 15197:2003 criteriaa R2 ISO 15197:2013 criteriab
CXT/CNEZ16,33 YSI 100% 100% 0.9964 100% in zone A
CNL13,33 YSI 100% 100% 0.9961 100% in zone Ac
CNUSB14,33 YSI 100% 100% 0.9957 100% in zone A
CN15,33 YSI 100% 99.8% 0.9959 100% in zone A
a

Within ISO 15197:2003 accuracy criteria (i.e., ≥95% of results shall fall within ±15 mg/dl or ±20% for samples with glucose concentrations < 75 and ≥75 mg/dl, respectively).

b

Within ISO 15197:2013 accuracy criteria (i.e., ≥95% of results shall fall within ±15 mg/dl or ±15% for samples with glucose concentrations <100 and ≥100 mg/dl, respectively).

c

Data on file.