Table 3. Qualitative Image Quality Parameters.

Cervicothoracic Spine Thoracolumbar Spine
Short-Bore MR Imaging Open MR Imaging P Value Short-Bore MR Imaging Open MR Imaging P Value
Arithmetic mean (SD) Arithmetic mean (SD)
T2w sagittal n = 18 n = 21 n = 30 n = 34
Overall image quality 2.1 (1.1) 3 (1) .006 1.8 (0.7) 2.7 (0.7) <.0001
Contrast 1.8 (0.9) 2.8 (0.8) .003 1.7 (0.7) 2.4 (0.7) <.0001
Contour sharpness 2.2 (1) 2.9 (0.9) .03 1.7 (0.7) 2.2 (0.6) .002
Noise 1.7 (0.6) 32 (0.8) <.0001 2.2 (0.6) 3.2 (0.7) <.0001
Artifactsa 2.6 (1) 2.2 (0.9) .25 2.1 (0.7) 2.3 (0.6) .11
T1w sagittal n = 18 n = 21 n = 30 n = 34
Overall image quality 1.8 (0.8) 3.5 (0.7) <.0001 1.8 (0.6) 2.7 (0.6) <.0001
Contrast 1.9 (0.8) 3.2 (0.7) <.0001 2 (0.7) 2.7 (0.6) <.0001
Contour sharpness 2.2 (0.9) 3.8 (0.5) <.0001 1.9 (0.6) 2.7 (0.6) <.0001
Noise 1.6 (0.6) 2.7 (1) .001 1.6 (0.5) 2.5 (0.6) <.0001
Artifactsa 1.7 (0.8) 3 (1) <.0001 1.6 (0.7) 2.6 (0.7) <.0001
T2w axial n = 19 n = 22 n  = 49 n = 54
Overall image quality 2.1 (0.7) 3.4 (0.7) <.0001 2 (0.7) 3.5 (0.6) <.0001
Contrast 2.2 (0.7) 3.2 (0.7) <.0001 2.1 (0.8) 3.3 (0.6) <.0001
Contour sharpness 2.1 (0.7) 3.2 (0.7) <.0001 1.8 (0.7) 3 (0.6) <.0001
Noise 1.8 (0.5) 3.5 (0.6) <.0001 2.2 (0.6) 3.9 (0.4) <.0001
Artifacts 2.7 (0.6) 2.7 (1) .9 2.1 (0.9) 3.1 (0.8) <.0001

An optimal score is defined as 1 and a poor score as 4. No sequence was rated nondiagnostic (score of 5) for any qualitative parameter.

a Artifacts were mainly due to motion for both, short-bore and open MR imaging.