Table 1.

Human case summary

# Type Detector Procedure Notes
Fluoro DSA
kV mA ms kV mA ms
1 Arteriovenous
malformation
Follow up.
II 71 50 13.3 85 320 35
MAF 71 50 13.3 85 160 25
2 Aneurysm
Follow up
Coil
II 96 50 13.3 85 320 93
MAF 90 50 10 86 160 41
MAF HD 80 160 20
80 160 6.3
3 Aneurysm
Initial
Coil + Stent
II 110 50 13.3 85 320 93
MAF 110 50 13.3 85 125 81
MAF HD 80 160 20
4 Aneurysm
Initial
Coil + Stent
II 110 50 13.3 85 320 35
MAF 90 50 10
MAF HD 80 160 20
5 Aneurysm
Initial
Coil
II 85 50 13.3 85 320 98
MAF 85 50 13.3 85 160 40
84 160 50
80 320 50 MFS
6 Aneurysm
Initial
Coil
II 100 50 13.3 85 320 35
MAF 100 50 13.3 85 160 25
7 Aneurysm
Initial
Coil
II 87 50 13.3
15hz
AP 97 320 85mfs
85 320 80
MAF 90 50 13.3 90 250 80
MAF HD 80 160 63
8 Aneurysm
Follow up
Coil
II 73 50 13.3 85 320 78 A
71 50 13.3 85 320 35 B
MAF 73 50 13.3 85 320 86 A
90 250 100 A
71 50 13.3 85 160 25 B
9 Aneurysm
Initial
Coil
II 77 50 13.3 97 125 77
MAF 90 50 10 80 160 25
10 Aneurysm
Follow up
Coil
II 76 50 13.3 110 100 100
MAF 99 50 13.3 90 160 20
11 Stenosis
Initial
Stent
II 110 49 13.3 85 320 35
MAF 110 49 13.3 80 200 6.3 MFS
80 125 25 SFS
15fps
12 Stenosis
Initial
Stent
II 88 50 13.3 85 320 88
MAF 90 42 11 80 200 20 SFS
15fps
13 Aneurysm
Initial
Coil
II 97 50 13.3 103 250 97
MAF 97 50 13.3
14 Aneurysm
Initial
Coil+ Stent
II 80 50 13.3 85 320 35
MAF 90 50 13.3 85 160 25
MAF HD 80 160 12
80 160 25
15 Stenosis
Initial
Balloon
Angioplasty
II 85 50 13.3 85 320 86 A
80 50 13.3 85 320 35 B
MAF 85 50 13.3 90 250 63 A
80 50 13.3 B
MAF HD 80 125 12 A
80 160 12 85 160 25 B
80 160 25 B

Notes:

1. For detector "HD" designates the MAF detector under high definition (HD) mode.

2. MFS refers to medium focal spot. SFS refers to small focal spot.

3. fps: frames per second.

4. The techniques used with the II for procedure segment A (or segment B) correspond to the techniques used with the MAF for segment A (or segment B). For example, for case #15, segment “A” for the II and the MAF used the same technique factors (85 kV, 50 mA, 13.3 ms).

5. AP, anterior posterior, refers to the Frontal II.

6. The technique parameters for MAF and II are selected to best reflect the response of the two detectors under similar conditions.