. 2013 Apr 9;17(1):191–198. doi: 10.1007/s11743-013-1469-x

Table 4.

Results of anionic surfactant determination by the standard MBAS method and MBAS continuous flow method for river water samples, calculated as SDBS equivalents (Warta River, Poznań)

Sample Standard MBAS method MBAS continuous flow method
Sample volume (mL) Absorbance Concentration (μg L−1) Sample volume (mL) Absorbance Concentration (μg L−1)
1 200 0.0273 55.0 100 0.1976 52.2
2 200 0.0250 50.0 100 0.1974 52.1
3 200 0.0200 39.0 100 0.1982 52.4
4 200 0.0249 49.9 100 0.1951 51.3
5 200 0.0242 48.3 100 0.1876 48.9
6 200 0.0244 48.8 100 0.2079 55.6
7 200 0.0253 50.7 100 0.1991 52.7
Mean 0.0244 48.8 0.1976 52.2
SD 0.0022 4.84 0.0060 1.98
RSD (%) 9.03 9.93 3.03 3.80