Measurements of force required to displace the catheter from the needle in Cannula P and I as an indirect measure of static resistance

  Cannula P Cannula I Difference p-value
22G static resistance; N
1.80 (25.2)
0.39 (28.7)
1.41 [1.09 to 1.73]
2 × 10-6
20G static resistance; N
0.73 (40.0)
0.55 (37.3)
0.19 [-0.04 to 0.41]
0.12
18G static resistance; N 2.40 (37.4) 0.44 (13.3) 1.96 [1.40 to 2.52] 7 × 10-5

Values are mean with (relative standard deviation %) or [95% confidence interval].