Table 4. Comparison between the reference method and PHE-Cambridge multiplex RT-PCR and HE-LAMP-BART.
Total | PHE RT-PCR Positive | PHE RT-PCR Negative | HE-LAMP-BART Positive | HE-LAMP-BART Negative | ||
PHE | Total | 218 | 106 | 112 | 106 | 112 |
Reference | Positive | 111 | 111 | 0 | 106 | 5 |
Test/method | Negative | 107 | 0 | 107 | 0 | 107 |
Sensitivity (95% CI) | 100% (96.73–100%) | 95.5% (89.8–98.52%) | ||||
Specificity (95% CI) | 100% (96.61–100%) | 100% (96.61–100%) |