Table 5.

Intertest variability of the probes of GeneXpert, calculated by standard deviation of the Cq values

GeneXpert probe and laboratory testing location Dilution Result with IS6110 copy no. of:
1 5 10 20
A Undiluted 1.2 2.0 2.8 1.1
    Lab 1 1:1,000 1.0 0.7 1.0 1.4
    Lab 2 1:1,000 NCa 2.7 1.5 8.2
    Lab 1b 1:10,000 0.9 0.8 2.8 1.8
    Lab 2 1:10,000 1.3 1.3 NDc 1.1
B Undiluted 1.4 2.1 2.9 1.6
    Lab 1 1:1,000 1.0 0.7 1.1 1.3
    Lab 2 1:1,000 NC 2.5 1.3 7.3
    Lab 1 1:10,000 1.0 0.7 2.3 1.9
    Lab 2 1:10,000 1.0 1.8 ND 1.0
C Undiluted 1.1 2.1 3.0 1.0
    Lab 1 1:1,000 1.0 0.7 1.1 1.4
    Lab 2 1:1,000 NC 2.7 1.4 8.3
    Lab 1 1:10,000 0.9 0.8 2.3 1.7
    Lab 2 1:10,000 1.3 1.6 ND 0.9
D Undiluted 1.1 1.7 2.8 1.0
    Lab 1 1:1,000 1.1 0.6 1.0 1.3
    Lab 2 1:1,000 NC 2.5 1.4 8.0
    Lab 1 1:10,000 0.9 0.8 2.0 1.7
    Lab 2 1:10,000 1.1 1.5 ND 1.1
E Undiluted 1.3 2.3 3.0 0.8
    Lab 1 1:1,000 1.1 0.7 1.0 1.3
    Lab 2 1:1,000 NC 2.8 1.6 ND
    Lab 1 1:10,000 0.9 0.9 2.3 1.8
    Lab 2 1:10,000 1.6 1.4 ND 1.2
a

NC, one of the three runs gave an error and thus the interest variability could not be calculated.

b

GeneXpert reached its analytical sensitivity with a dilution of 1:10,000 in lab 1.

c

ND, at least one run could not be detected; thus, the analytical sensitivity was not reached at this dilution.