. 2013 Spring;21(1):37–40. doi: 10.1177/229255031302100109

TABLE 1.

Mean radiation dose measured from the vicinity of the metallic port of the tissue expander, and corresponding locations over the non-port-containing permanent saline implant

Location Dose around, cGy
P*
Tissue expander with port Saline implant without port
6 MV photons
  Over port 208.2 210.4 0.61
  Medial/lateral to port 205.7 206.6 0.86
  Inferior to port 204.9 205.3 0.97
18 MV photons
  Over port 211.0 205.9 0.33
  Medial/lateral to port 196.3 200.6 0.40
  Inferior to port 196.0 194.8 0.69
9 MeV electrons
  Over port 190.6 188.3 0.57
  Adjacent to port 171.4 168.9 0.70
12 MeV electrons
  Over port 190.1 188.5 0.64
  Adjacent to port 187.1 178.8 0.15
*

Two-sided t test