. 2013 Nov-Dec;21(6):505–510. doi: 10.1590/1679-775720130083

Table 2.

Mean (SD) of the number of cycles to failure (NCF) and fragment lengths in mm (SD) of tested NiTi instruments according to the groups

NiTi instruments Groups NCF (SD) Fragment length mm (SD)
ProFile (n=40) Control 396.5 (77.46) 2.6 (0.4)
5 min %2.5 NaOCl 385.5 (76.03) 2.4 (0.5)
5 min %2.5 NaOCl + sterilization 359 (73.25) 2.5 (0.6)
5X (5 min %2.5 NaOCl + sterilization) 320 (61.55) 2.6 (0.2)
FlexMaster (n=40) Control 379 (70.54) 3 (0.4)
5 min %2.5 NaOCl 378 (53.39) 2.5 (0.4)
5 min %2.5 NaOCl + sterilization 374 (70.82) 2.7 (0.3)
5X (5 min %2.5 NaOCl + sterilization) 373.5 (67.37) 2.5 (0.4)
Twisted File (n=40) Control 551.2 (118.78) 2.4 (0.4)
5 min %2.5 NaOCl 526.5 (105.8) 2.4 (0.2)
5 min %2.5 NaOCl + sterilization 439 (148.16) 2.6 (0.4)
5X (5 min %2.5 NaOCl + sterilization) 415.7 (126.21) 2.5 (0.2)
Mtwo (n=40) Control 599.5 (79.46) 2.8 (0.2)
5 min %2.5 NaOCl 560 (56.81) 2.7 (0.2)
5 min %2.5 NaOCl + sterilization 553 (87.94) 2.6 (0.2)
5X (5 min %2.5 NaOCl + sterilization) 514.5 (73.76) 2.6 (0.2)

SD=Standard deviation