Table 1. Comparison of Commercial Handheld OCT Systems to the Prototype Handheld OCT System.

Specification Bioptigen Envisu Optovue iVue Prototype Handheld
OCT Technology SD-OCT SD-OCT SS-OCT
Imaging Speed 32,000 A-scan/s 26,000 A-scan/s 350,000 A-scan/s
Imaging Range in Tissue 1.7 mm (C2200)
2.5 mm C2300)
2 – 2.3 mm 3.08 mm
Axial Resolution in Tissue 6 µm (HR source)
3 µm (VHR source)
5 µm 10 µm
Operating Wavelength ~840 nm ~840 nm ~1060 nm
Scanning Mirror Type Galvanometer Galvanometer 2D MEMS
Dimensions
(L x H x W)
23 cm x 18 cm x 8 cm
(including lens)
28 cm x 32 cm x 14 cm 15 cm x 18 cm x 7 cm (Power Grip)
16 cm x 10 cm x 14 cm
(Camcorder)
Handheld Weight 1.5 kg (with lens) 2.2 kg 0.50 kg (Power Grip)
0.42 kg (Camcorder)
Refractive Error Correction Manual Motorized Manual
Display on Handheld N/A LCD Touch Screen LCD Screen
Alignment Aids OCT Preview Scans OCT Preview Scans
IR Camera
Fixation Target
OCT Preview Scans
Iris Camera
Fixation Target
Ophthalmic Imaging Retinal
Anterior Segment
Retinal
Anterior Segment
Retinal