Table 1.

Mean ± standard deviation (SD) errors in photopic illuminance measurements, relative to a calibrated meter (Gigahertz-Optik), from six Daysimeter (Daysimeter-D) and six Actiwatch Spectrum devices under diffuse illumination from six different practical light sources.

Source Calibrated light meter (lux) Daysimeter (lux) Daysimeter measurement error (%) Daysimeter measurement range (lux) Actiwatch (lux) Actiwatch measurement error (%) Actiwatch measurement range (lux)
Philips 427 430±5.83 0.65±1.37 422 – 440 (range = 4%) 585±20.5 37.1±4.80 564 – 619 (range = 9%)
F32/T8/TL/735 (3500 K fluorescent)

GE 193 191±2.37 −1.21±1.22 188 – 195 (range = 3%) 332±13.5 71.9±6.99 309 – 349 (range =11%)
F55BX/AR/FS (9000 K fluorescent)

Sylvania LU 250 (high pressure sodium) 103 96.4±2.49 −6.70±2.41 92.1 – 99.0 (range = 7%) 40.2±1.54 −61.1±1.49 38.3 – 48.3 (range = 20%)

Daylight 1944 1950±31.4 0.29±1.61 1920 – 1991 (range = 4%) 3886±160 99.9±8.20 3649 – 4141 (range = 12%)

Beta Edge 5760 K 75.4 71.5±0.89 −5.27±1.18 70.2 – 72.5 (range = 3%) 81.9±6.39 8.54±8.48 73.4 – 89.7 (range =18%)
LED