Skip to main content
. 2013 Dec 20;11:39. doi: 10.1186/2052-336X-11-39

Table 1.

The instrumental settings for the ICP-MS

Spray chamber:
Quartz Cyclonic
Nebulizer:
Meinhard® Concentric Quartz Type A0.5
Torch:
Standard Quartz
Sampler Cone:
Platinum
Skimmer Cone:
Platinum
Plasma gas flow:
18 L min-1
Auxiliary gas flow:
1.20 L min-1
Nebulizer gas flow:
1.04 L min-1
ICP RF power:
1550 W
Pulse voltage:
900 V
Analog stage voltage:
−1637 V
KED mode cell entrance voltage:
−4 V
KED mode cell exit voltage:
−38 V
KED mode axial field voltage: 475 V