Published in final edited form as: Invest Radiol. 2013 Jun;48(6):437–444. doi: 10.1097/RLI.0b013e31828027c2

Table 1.

Specific parameters for the acquisitions

Acquisition Planning Targeting* Target Verification Monitoring Verification

Sequence bSSFP
(‘trueFISP’)
bSSFP
(‘BEAT_IRTTT’)
bSSFP
(‘trueFISP’)
HASTE HASTE 3D FSPGR (‘VIBE’)

TR (ms) <4 <4 <4 2000 2000 6

TE (ms) <3 <3 <3 83 83 2.8

Flip Angle 55–70 55–70 55–70 ** ** 15

Matrix 256×192 192×192 256×192 192×192 192×192 256×134***

FOV (cm) 20–36 20–36 20–36 20–36 20–36 35×26

Slice/Skip (mm/mm) 5/0 4/1 5/0 5/2.5

BW (Hz/pix.) 500–800 500–800 500–800 400–500 400–500 250

Slices 15–25 3 15–25 52

Acq. Time (s) < 1sec/img <0.8 sec/img <0.8 sec/img 2 sec/img 2 sec/img <20 sec/acq

*

Prototype interactive bSSFP pulse sequence, Siemens Corporate Research, sequence author: Li Pan

**

Standard 90° excitation with 150° refocusing pulses

***

Parallel imaging acquisition and reconstruction (GRAPPA) with 75% phase field of view and 70% sampling in slice direction.