Published in final edited form as: J Magn Reson Imaging. 2012 Nov 29;38(1):217–224. doi: 10.1002/jmri.23875

Table 1.

T1 Phantom Measurements for MOLLI Pulse Sequences at HR of 60 or 100 BPM

Reference 3-3-5 HR 60 3-5 HR 60 2-2-4 HR 60 3-3-5 HR 100 3-5 HR100 2-2-4 HR 100
1459.1 1487 (1.8%) 1492 (2.2%) 1384 (5.1%) 1345 (7.8%) 1373 (5.8%) 975 (33%)
1150.3 1088 (5.5%) 1084 (5.7%) 1054 (8.4%) 1033 (10%) 1042 (9.5%) 925 (20%)
855.6 819 (4.3%) 809 (5.4%) 807 (5.7%) 801 (6.4%) 797 (6.8%) 756 (12%)
699.3 680 (2.7%) 674 (3.7%) 678 (3.1%) 672 (3.9%) 669 (4.4%) 651 (6.7%)
630.9 629 (2.7%) 626 (0.8%) 629 (3.3%) 624 (1.1%) 622 (1.4%) 611 (3.2%)
568.4 538 (5.3%) 535 (5.9%) 539 (5.2%) 535 (5.8%) 534 (6.1%) 531 (6.7%)
521.1 483 (7.1%) 479 (8.0%) 484 (7.1%) 480 (7.8%) 476 (8.7%) 476 (8.5%)
479.5 447 (6.7%) 444 (7.5%) 448 (6.6%) 444 (7.4%) 441 (6.6%) 443 (7.7%)
447.0 422 (5.4%) 420 (6.1%) 423 (5.5%) 419 (6.2%) 417 (5.0%) 419 (6.2%)

Percent error from IR gradient echo reference in parentheses.