. 2013 Oct 15;23(3):308–315. doi: 10.11613/BM.2013.037

TABLE 2.

Improvement of phlebotomy error rates pre- and post- training of H03-A6 and proposed new procedure.

Error description Laboratories before training# Laboratories after training with CLSI H03-A6 document## Laboratories after training with new procedure

All (N = 30) Public Lab (N = 15) Private Lab (N = 15) P All (N = 30) Public Lab (N = 15) Private Lab (N = 15) P All (N = 30) Public Lab (N = 15) Private Lab (N = 15) P
Inappropriate request to the patient to clench the fist repeatedly 25/30 14/15 11/15 0.329 29/30* 15/15 14/15 1.000 0/30*** 0/15 0/15 ---
Inadequate friction procedure during the cleaning of the venipuncture site 27/30 13/15 14/15 1.000 0/30** 0/15 0/15 --- 0/30 0/15 0/15 ----
Incorrect sequence of vacuum tubes 26/30 13/15 12/15 1.000 0/30** 0/15 0/15 --- 0/30 0/15 0/15 ----
Incorrect mixing of vacuum tubes 25/30 15/15 10/15 0.042 0/30** 0/15 0/15 --- 0/30 0/15 0/15 ----

Comparison of error rates between public and private laboratories before training (McNemar Chi-square test,

*

P = 0.113 and

**

P < 0.001), and after training to the proposed new procedure (Fisher exact test two-tailed test

***

P < 0.001

****

P = 0.237). ---, not calculated;

#

data previously published (2);

##

data previously published (4).