Skip to main content
. 2014 Jan 15;16(1):e11. doi: 10.2196/jmir.2798

Table 1.

Mean difference and standard deviation of the difference with corresponding confidence intervals at P=.05 of hearing thresholds and calibration coefficients between series 1 and 3 calculated jointly for 8 frequencies (N=25).

Method Difference (dB),
mean (95% CI)
Difference (dB),
SD (95% CI)
Hearing threshold


Right ear –0.38 (–1.13, 0.38) 5.40 (4.92, 5.99)

Left ear 0.13 (–0.62, 0.87) 5.34 (4.86, 5.92)

Both ears –0.13 (–0.65, 0.40) 5.37 (5.02, 5.77)

Bilateral –0.35 (–1.04, 0.34) 4.92 (4.48, 5.46)
Calibration coefficient


Modulated signal –0.09 (–0.78, 0.61) 4.97 (4.53, 5.51)

Dual tone (5 dB) 1.05 (0.00, 2.11) 7.54 (6.87, 8.37)

Dual tone (2 dB) 1.20 (0.06, 2.35) 8.18 (7.45, 9.07)

Ascending (5-dB step) –0.15 (–1.00, 0.70) 6.05 (5.51, 6.71)

Ascending (2-dB step) –0.10 (–0.80, 0.60) 5.00 (4.55, 5.54)

Bekesy (continual) 0.88 (0.19, 1.57) 4.92 (4.48, 5.46)

Bekesy (modulated) 0.63 (0.09, 1.17) 3.87 (3.52, 4.29)