TABLE 1.

Mean and SD of 10 Measurements of Residual Activity in 6 Different Types of Syringes

Syringe Syringe capacity (mL) Manufacturer Brand Residual activity (%)
Mean SD P
Brand A 3 mL Covidien Monoject 3 cc 19.7 2.6
Brand B 3 mL BD Medical BD 3 cc 21.5 5.6 NS
Brand C 1 mL Covidien Monoject 1 cc insulin 17.5 3.2 NS
Brand D 1 mL Covidien Monoject 1 cc tuberculin 15.6 2.8 P < 0.01
Brand E 1 mL Terumo Medical Terumo 1 cc 13.1 3.0 P < 0.001
Brand F 3 mL DPS Meditech DPS 3 cc 11.4 3.6 P < 0.001