Exposure indicator DI control limits for clinical images.

DI Range Action
>+3.0 Excessive patient radiation exposure: Repeat only if relevant anatomy is “burned out,” require immediate management follow-up
+1.0 to +3.0 Overexposure: Repeat only if relevant anatomy is burned out
−0.5 to +0.5 Target range
Less than −1.0 Underexposed: Consult radiologist for repeat
Less than −3.0 Repeat