. 2013 Oct 31;31(1):25–33. doi: 10.1007/s10815-013-0107-x

Table 3.

From experiment 2–means and standard deviations (SD) of a second lot of Accu-Beads with the use of the MicroCell counting chamber and the Standard Count counting chamber

Item Method N Mean ± SD (M/mL) P value
MicroCell Manual 30 21.1 ± 1.3 0.97
CASA 30 21.1 ± 1.3
Manual 30 41.5 ± 1.9 0.67
CASA 30 41.7 ± 1.6
Standard count Manual 30 17.9 ± 1.9 0.32
CASA 30 17.5 ± 1.2
Manual 30 37.4 ± 3.6 0.44
CASA 30 36.9 ± 2.0

Samples counted manually and with a computer automated semen analyzer (CASA)