. 2014 Feb;52(2):517–523. doi: 10.1128/JCM.02461-13

TABLE 3.

Discordance for HIV-1 RNA levels at <200 copies/mla

Assay (at <200 copies/ml) % discordant results of ≥200 copies/ml (no. of discordant samples/total no. of samples)
Amplicor RealTime TaqMan v1 TaqMan v2
Amplicor NA 5 (58/1,161) 3 (7/204)
RealTime NA 10 (37/356) 7 (49/728)
TaqMan v1 7 (85/1,188) 16 (59/378) 7 (25/236)
TaqMan v2 3 (7/204) 3 (22/701) 1 (3/314)
a

Shown are the percentages of results that were discordant between the assays at a threshold of 200 copies/ml. The percentage of discordant samples decreased for all comparisons when a threshold of 200 copies/ml was used, compared to a 50-copies/ml threshold. NA, not available.