Skip to main content
. 2014 Jan;52(1):103–107. doi: 10.1128/JCM.02460-13

TABLE 1.

Specificities and sensitivities of the BioPlex 2200 HIV Ag-Ab assay

Screening and result No. of samples that were:
Sensitivity (% [95% CI]) Specificity (% [95% CI])
HIV infected (n = 35) HIV uninfected (n = 1,470)
Initial
    Positive 35 9
    Negative 0 1,461
    Performance 100 (96.7–100) 99.4 (99.0–99.8)
Retest
    Positive 35 8
    Negative 0 1,462
    Performance 100 (96.7–100) 99.5 (99.1–99.9)