Table 3.

Interference from Common Hemoglobin Variants on A1c Measurements from Commonly Used Methods24

Method Interference from HbC Interference from HbS Interference from HbE Interference from HbD
Abbott Architect/Aeroset Yes Yes * *
Arkray ADAMS A1c HA-8180V (Menarini) No No A1c not quantified A1c not quantified
Axis-Shield Afinion No No No No
Bayer A1cNOW Yes Yes No No
Beckman AU system Yes Yes No No
Beckman Synchron System No No No No
Bio-Rad D-10 (A1c program) No No No No
Bio-Rad Variant II NU No No
Bio-Rad Variant II Turbo No No Yes Yes
Bio-Rad Variant II Turbo 2.0 No No No/Yes (conflicting reports) No
Bio-Rad in2it Yes No Yes No
Ortho-Clinical Vitros No No No No
Roche Cobas Integra Gen.2 No No No No
Roche/Hitachi (Tina Quant II) No No No No
Sebia Capillarys 2 Flex Piercing No No No No
Siemens Advia HbA1c (original version) Yes Yes * *
Siemens Advia A1c (new version) No No * *
Siemens DCA 2000 No No No No
Siemens Dimension No No No No
Tosoh G7 Yes No Yes No
Tosoh G8 No No Yes No
Trinity (Primus) HPLC (affinity) No No No No

* Generally assumed there is no interference

Adapted with permission from: HbA1c assay interferences. National Glycohemoglobin Standardization Program Web site. Available at: http://www.ngsp.org/interf.asp. Updated August, 2012. Accessed November 8, 2012