Skip to main content
. 2014 Jan 22;4:1. doi: 10.1186/2045-9912-4-1

Table 3.

Effects of pulse time and pulse delay on nitric oxide uptake efficiency for varied breathing patterns

 
 
Uptake efficiency
Pulse
Pulse
Vt = 500 ml
Vt = 500 ml
Vt = 250 ml
Vt = 250 ml
Time [ms] Delay [ms] f = 12 min -1 f = 25 min -1 f = 12 min -1 f = 25 min -1
100
0
0.94
0.89
0.88
0.81
 
50
0.94
0.89
0.85
0.72
 
100
0.94
0.89
0.81
0.59
 
200
0.94
0.88
0.70
0.24
200
0
0.94
0.89
0.84
0.70
 
50
0.94
0.89
0.80
0.57
 
100
0.94
0.88
0.76
0.42
 
200
0.94
0.86
0.63
0.14
400
0
0.94
0.88
0.74
0.42
 
50
0.94
0.86
0.68
0.31
 
100
0.94
0.81
0.61
0.22
  200 0.93 0.62 0.46 0.07