TABLE 4.

Estimated Eye Lens Dose to the Patient From Brain Perfusion CT

Scanner Tube Potential Setting (kVp)
80 100 120 140

Sensation 64 2.4 5.2  8.9 14.5
VCT 4.1 7.1 10.7 14.7
Brilliance 64 2.2 NA  6.7 10.4
Aquilion 64 4.4 7.7 11.5 14.7a

Note—The eye lens dose is normalized to the unit of mGy/100 mAs. Sensation 64 manufactured by Siemens Healthcare, VCT manufactured by GE Healthcare, Brilliance 64 manufactured by Philips Healthcare, and Aquilion 64 manufactured by Toshiba. NA = not applicable.

a

Aquilion 64 provides a tube potential setting of 135 instead of 140 kVp.