Abstract
Introduction: Conventional formalin-fixed, paraffin-embedded tissue provides superior cellular morphology and long-term storage. Problems with formalin fixation comprise delay of fixation and variations in the duration of fixation. Microwave assisted tissue fixation removes the use of noxious and potentially toxic formalin that decreases the turnaround time and creates a personnel friendly workflow.
Material and Methods: The present study was conducted over a period of two years. One hundred and forty paired tissue sections were taken including both neoplastic and non-neoplastic tissues. One of the paired tissues was fixed in formalin and the other was fixed by using microwave irradiation in phosphate buffered saline. Both were then processed by conventional method. Each slide was examined and rated for the adequacy of fixation by two pathologists in a blinded fashion using 7 parameters: Cellular outline, cytoplasmic detail, nuclear detail, erythrocyte integrity, lymphocyte appearance, overall morphology and overall staining.
Results: Statistical analysis showed that sections obtained from microwave fixed tissues were comparable to that of routinely fixed tissue. The p-values of all parameters were not significant except for the overall morphology for which p-value was significant owing to loss of tissue in some cases.
Conclusion: Microwave irradiation substantially shortened the time from specimen reception to diagnosis (turnaround time) and allowed same-day tissue processing and diagnosis of specimens without compromising the overall quality of the histologic section.
Keywords: Microwave, Fixation, Turnaround time
Introduction
Most methods currently used to manipulate tissues for microscopic examination were developed in the early 1900s. These methods evolved largely after the establishment of formalin as a fixative that could rapidly and permanently preserve fairly large tissue specimens [1].
Fixation is the first step in tissue preservation for pathological diagnosis. Fixation arrests autolysis and putrefaction, coagulates soluble and structural proteins, fortifies tissues against the deleterious effects of subsequent processing, and facilitates staining. Fixation of tissues can be accomplished by formalin to form cross-links in tissues. However, formalin fixation has three disadvantages:
Slow fixation (16–24 hours) hinders intraoperative decision making.
Slow quenching of enzymatic activity causes RNA degradation.
Extensive molecule modification affects protein antigenicity [2].
Changing the standard technique for tissue fixation and preparation from the currently used overnight processing to same day microwave tissue preparation could substantially reduce turnaround times, permitting same-day diagnosis that would facilitate patient diagnosis and management on a one day basis. This improvement in turnaround time could reduce costs associated with diagnosis and patient dissatisfaction. It increases the rapidity with which neoplastic diseases are diagnosed and therapy is initiated [3,4].
‘Microwaves’ are defined as electromagnetic waves with a frequency between 300MHz and 300GHz. The advantage of microwave use is decreased fixation time but disadvantages include tissue shrinkage and breakdown of RBCs, however, these artefacts are very minor if the technique is carried out carefully so that no appreciable differences with routinely processed material is evident [5].
The microwave method shortened the total time of fixation and dehydration required by the conventional method [6].
For almost 100 years, the steps followed to prepare tissues for microscopic review have remained practically unchanged. [7] Formaldehyde-Fixed Paraffin-Embedded tissue (FFPE) is the product of a century old histopathology practice. However, tissue processed by this system has limited application beyond routine histology and immunohistochemistry [8]. Substantial shortcomings associated with this practice include at least a one day delay in providing the diagnosis, [9,10] reagent toxicity, and degradation of nucleic acids [11].
In present study routine and rapid methods of tissue fixation were compared in terms of quality of histological sections and turnaround times.
Aims and Objectives
To compare routine and rapid tissue fixation in terms of quality and reliability of histological sections.
To assess the time factor of rapid tissue fixation as against routine processing.
Material and Methods
This was a prospective study conducted over a period of two years (August 2010-July 2012) on the tissue specimens received from surgical departments in Subharti Medical College and associated chhatrapati Shivaji Hospital, Meerut, India.
A pilot study was done on 10 randomly selected tissue specimens using a domestic microwave oven. Alterations in the time for fixation, power (watt) and Variation in the pH of buffer was done. Tissues fixed at low power for less time showed suboptimal fixation whereas at high power and more time tissue destruction was noted. Change in pH of the buffer however caused no significant difference. Since an accurate temperature could not be reached, results of fixation were unsatisfactory.
Hence, an Enzyme retriever microwave, equipped with temperature probe accurate to ± 1oC was used thereafter and in the present study only cases processed by latter technique were analysed.
The material for the study constituted 140 paired tissue specimens taken randomly (including both neoplastic and non-neoplastic tissues). Very tiny endoscopic biopsies were excluded eg: skin biopsies or mucosal biopsies. Tissues were received in 10% Formalin as a transport or collection medium. Two slices not larger than 5 mm in thickness were selectively taken from each gross specimen received. One was kept overnight in formalin and processed conventionally.
The other slices were placed in plastic cassettes. Four to five cassettes were then immersed in approximately 500ml PBS 0.1M pH 7.4 in a microwave-safe container and microwaved at 68°C for 5 minutes in enzyme retreiver V.2.2 microwave equipped with a temperature probe accurate to +/- 1°C [Table/Fig-1, 2]. Conventional histoprocessing and staining were done. Automated tissue processor Shandon Citadel 2000 was used for histoprocessing. After staining the slides, each slide (of both tissues fixed in formalin and by microwave irradiation) was examined and rated for the adequacy of fixation by two pathologists in a blinded fashion, applying a scale of 1 to 5 (1=poor, 2-fair, 3=good, 4=excellent, 5=outstanding) [12]. Seven parameters were used: Cellular outline, cytoplasmic detail, nuclear detail, erythrocyte integrity, lymphocyte appearance, overall morphology and overall staining. Overall grading was done by adding the scores of each parameter:- 7-13 –>Poor, 14-20–>Fair, 21-27–>Good, 28-34–>Excellent, >34–>Outstanding.
Data management and Statistical analysis was performed using SPSS software version 20.0. Chi-square test was used. The p-value < 0.05 was considered statistically significant.
Results
A total of 140 paired tissue samples were fixed by microwave irradiation and conventional method and comparison of the quality of histologic section and turnaround time was determined [Table/Fig-3].
After fixation by microwave irradiation, the tissue sections were firm, shrunken and pale. On microscopic examination by two pathologists, the inter-observer variation was found to be minimal. The cellular outline, cytoplasmic detail, nuclear detail, erythrocyte integrity, lymphocyte appearance and staining as seen after rapid fixation were comparable to that of routine fixation. The p-values were not significant [Table/Fig-4, 5a,B,C,D,E,F,g].
There was partial loss of tissues in rapidly fixed tissues owing to effacement of tissue architecture hence the overall morphology was graded slightly inferior to that of routinely fixed tissues. p value was 0.0001 which is statistically significant [Table/Fig-6].
Out of the 140 paired tissues processed, 20 tissues fixed by rapid method were graded as ‘EXCELLENT’ in comparison to 43 of routine fixation. One hundred and eleven cases fixed by rapid method were graded as ‘GOOD’ in comparison to 92 fixed by conventional formalin fixation. 8 rapidly fixed tissues were graded ‘FAIR’ while only 4 were graded so by conventional method. One tissue each was graded as ‘POOR’ on fixation by both rapid and routine method. The p-value of 0.0002 was significant [Table/Fig-7].
Better morphological details in microwave fixed tissues were noted in all cases of adipose tissue, placental tissue and small specimens like endometrial biopsy and prostate chips. Staining was also sharper in these tissues when compared with formalin fixed tissues.
Tissues were rapidly fixed in 5 mins by microwave irradiation whereas formalin fixation took almost 12 hours. The slides of the tissues fixed by rapid microwave irradiation were received within 24 hours of receiving the specimen when processed by the automated tissue processor whereas the turnaround time was a minimum of two days for conventionally fixed tissues
Discussion
Microwaves are non ionizing radiations that produce alternating electromagnetic fields that result in rotation of dipolar molecules such as water and the polar side chains of proteins through 1800 at the rate of 2.45 billion cycles/second. The molecular kinetics so induced results in the generation of energy flux and continues till radiation ceases [12].
In the present study microwave irradiation has been used only for fixation of tissues like Mayer CP [13] and Pitol DL et al.,[14].
In this study phosphate buffered saline was used as a medium for microwave irradiation and tissues were irradiated at 680C. Login GR et al., [15] found that tissues irradiated to 50± 50C were superior to those at low temperature. Many other studies showed that use of normal saline gave better fixation results than using formalin in microwave irradiation eliminating the noxious and potentially toxic fumes [14,16,17]. Choudhary K et al., [18] demonstrated that only 5 out of 16 tissue (30%) were stabilized in normal saline whereas 22 out of 34 (80%) fixed in formalin with microwave exposure showed satisfactory results.
After fixation by microwave the tissues became firm and pale in colour as also noted by Mayer CP [13] and Leong ASY et al., [17] on macroscopic examination.
On microscopic examination tissue was partially lost in 39.28% cases, predominantly in tissues like gall bladder, lymphoid tissues, cervix and tumour tissues. However, Morales et al., [19] found no loss of tissue architecture between conventionally and microwave processed tissue.
Ragazzini T et al., [20] observed that the loss of tissue in domestic microwave was due to the fact that a constant temperature cannot be reached or maintained in a specific time frame which has now been overcome by computerized systems. In other studies this parameter has not been used as a part of their analysis.
The loss of tissue in this study was seen maximum in cases of lymphoid tissue. All lymphoid tissues were received in fresh state and were not partially fixed. Tissues kept in formalin for 1-2 hours before microwave irradiation gave the best results without any loss of architectural details. In another study, it was found that fixation in neutral buffered formalin followed by microwave irradiation gave better results when compared to microwave irradiation alone [21]. Similar finding was obtained on microwave fixation of tissues previously soaked for four hours in formaldehyde [22].
The cellular outline and cytoplasmic details of the microwave fixed tissue was comparable with that of conventional formalin fixed tissues. This is consistent with other studies [12,23]. However, in our study the nuclear details were slightly less clear in microwave fixed tissues. This finding is supported by the study of Panja P et al., [24] who found that there was loss of nuclear details under high power when complete histoprocessing was done using the microwave oven whereas Kennedy A and Foulis AK [25] and Reed W et al., [26] found that the cytological details of the nuclei was much clearer using the microwave technique.
In the microwave irradiated tissues there was no loss of erythrocytes which is in contrast to the studies of Mayer CP [13] and Hopwood D et al., [27] who demonstrated erythrocytolysis. Boon ME et al., [22] performed the microwave irradiation of brain and found that the erythrocytes were intact. In both methods of fixation inflammatory cells like plasma cells and lymphoctyes were distinguishable from one another as is also demonstrated in this study.
The overall morphology was slightly inferior to that of conventionally formalin fixed tissues owing to the partial loss of tissue which is not in agreement with other studies [1,13,28–30].
Staining of the rapid microwave fixed tissues was sharper and brighter in most of the tissues than the tissues obtained after conventional fixation. This is consistent with the findings of Panja P et al., [24], Barrett C et al., [31] and Ragazzini T et al., [20] who showed that the microwave processed tissue showed a brighter staining with eosin and a stronger reaction with haematoxylin as compared to routinely processed tissue.
Neoplastic tissues in the present study were equally well preserved as the non-neoplastic tissues in the process of microwave fixation. This was similar to the findings of Hopwood D et al., [27] and Emerson LL et al., [32] who showed that malignant cells were equally well preserved by both conventional and microwave fixed tissues. Right from precancerous conditions and lesions to malignancy, reactive lesions to benign tumours, microwave technique has shown a remarkable difference without losing the architecture and morphology of the cells [23].
In this study adipose tissue showed the best morphology and staining characters post microwave irradiation. This is in agreement with Babu TM et al., [4] Other observers like Suri V et al., [33] found that tissues containing fat responded poorly to microwave irradiation. Steri V et al., [34] also found that adipose tissue showed artifacts due to incomplete fixation.
Small specimens like endometrial currettings and prostate chips were morphologically better when fixed in buffered saline in microwave oven. This is in agreement with the findings of Hafajee ZAM et al., [35] and Rohr LF et al., [3].
The overall time for fixation was reduced from 12 hours to 5 minutes when using microwave irradiation. This was found in many studies where fixation time was decreased from hours to a few minutes [17,18,27,30,36,37].
Conclusion
This study demonstrates that microwave irradiation under controlled conditions can reduce fixation time to about 5 minutes and thus reduce the overall processing time in surgical pathology without significant loss of specimen quality. This allowed same-day tissue processing and diagnosis of specimens which is essential when urgent reporting is required. It eliminates the use of formalin which is a potential hazard for health.
Adoption of microwave fixation in the regular histopathology laboratory on a regular basis can only be done after standardizing the time and temperature for each specimen as different tissues respond differently to microwave irradiation. Thus drawbacks like partial loss of tissues could be minimized.
Financial or Other Competing Interests
None.
References
- [1].Cox ML, Schray CL, Luster CN, Stewart ZS, Korytko PJ, Khan KNM, et al. Assessment of fixatives, fixation and tissue processing on morphology and RNA integrity. Experimental and Molecular Pathology. 2006;80:183–91. doi: 10.1016/j.yexmp.2005.10.002. [DOI] [PubMed] [Google Scholar]
- [2].Chu W, Furusato B, Wong K, Sesterhenn IA, Mostofi FK, Wei MQ, et al. Ultrasound accelerated formalin fixation of tissue improves morphology, antigen and mRNA preservation. Modern Pathology. 2005;18:850–63. doi: 10.1038/modpathol.3800354. [DOI] [PubMed] [Google Scholar]
- [3].Rohr LR, Layfield LJ, Wallin D, Hardy D. A Comparison of Routine and Rapid Microwave Tissue Processing in a Surgical Pathology Laboratory. Am J Clin Pathol. 2001;115:703–8. doi: 10.1309/15fb-fld1-408x-jqa3. [DOI] [PubMed] [Google Scholar]
- [4].Babu TM, Malathai N, Magesh KT. A comparative study on microwave and routine tissue processing. Indian J Dent Res. 2011;22:50–5. doi: 10.4103/0970-9290.79975. [DOI] [PubMed] [Google Scholar]
- [5].Rosai J. Rosai & Ackerman’s Surgical pathology. 10th ed. India: Elsevier; 2009. Gross techniques in surgical pathology. In: Damjanov I, Rosai J, Nola M, editors; p. 27-9. [Google Scholar]
- [6].Wagenaar F, Kok GL, Davies JMB, Pol JMA. Rapid cold fixation of tissue samples by microwave irradiation for use in electron microscopy. The Histochem Journal. 1998;25:719–25. doi: 10.1007/BF00211767. [DOI] [PubMed] [Google Scholar]
- [7].Morales AR, Nassiri M, Kanhoush R, Vincek V, Nadji M. Experience with an automated microwave-assisted rapid tissue processing Method. Validation of Histologic Quality and Impact on the Timeliness of Diagnostic Surgical Pathology. Am J Clin Pathol. 2004;121:528–36. doi: 10.1309/ACK8-AHV0-1T47-QR53. [DOI] [PubMed] [Google Scholar]
- [8].Nassiri M, Ramos S, Zohourian H, Vincek V, Morales AR, Nadji M. Preservation of biomolecule in breast cancer by a formalin free histology system. BMC Clinical Pathol. 2008;8:1045–56. doi: 10.1186/1472-6890-8-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [9].Zarbo RJ, Gephardt GN, Howanitz PJ. Intra-laboratory timeliness of surgical pathology reports. Arch Pathol Lab Med. 1996;120:234–44. [PubMed] [Google Scholar]
- [10].Novis DA, Zarbo RJ, Saladino AJ. Interinstitutional comparison of surgical biopsy diagnosis turnaround time. Arch Pathol Lab Med. 1998;122:951–6. [PubMed] [Google Scholar]
- [11].Lewis F, Maughan NJ, Smith V, Hillan K, Quirke P. Unlocking the archive: gene expression in paraffin-embedded tissue. J Pathol. 2001;195:66–71. doi: 10.1002/1096-9896(200109)195:1<66::AID-PATH921>3.0.CO;2-F. [DOI] [PubMed] [Google Scholar]
- [12].Mathai AM, Naik R, Pai MR, Rai S, Baliga P. Microwave histoprocessing versus conventional histoprocessing. Indian J Pathol Microbiol. 2008;51:12–6. doi: 10.4103/0377-4929.40383. [DOI] [PubMed] [Google Scholar]
- [13].Mayers CP. Histological fixation by microwave heating. J Clin Pathol. 1970;23:273–5. doi: 10.1136/jcp.23.3.273. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [14].Pitol DL, Issa JPM, Camacho E, Wolga ND, Caetano FH, Lunardi LO. Int J Morphol. 2007;25(4):695–701. [Google Scholar]
- [15].Login GR, Dvorak AM. Microwave energy fixation for electron microscopy. Am J Pathol. 1985;120:230–43. [PMC free article] [PubMed] [Google Scholar]
- [16].Login GR. Microwave fixation versus formalin fixation of surgical and autopsy tissues. American J med technol. 1978;44:435–7. [PubMed] [Google Scholar]
- [17].Leong ASY, Duncis CG. A method of rapid fixation of large biopsy specimens using microwave irradiation. Pathology. 1986;18:222–5. doi: 10.3109/00313028609059463. [DOI] [PubMed] [Google Scholar]
- [18].Chaudhary K, Chattopadhyay A, Dutta SK. Microwave technique in histopathology and its comparison with the conventional technique. Indian J Pathol Microbiol. 2000;43(4):387–94. [PubMed] [Google Scholar]
- [19].Morales AR, Essenfeld H, Essenfeld E, Duboue MC, Vincek V, Nadji M. Continuous-specimen-flow, high-throughput, 1-Hour tissue processing. Archives Pathol Lab Med. 2002;126:583–90. doi: 10.5858/2002-126-0583-CSFHTH. [DOI] [PubMed] [Google Scholar]
- [20].Ragazzini T, Magrini E, Cucchi MC, Foschini MP, Eusebi V. The Fast Track Biopsy (FTB): Description of a rapid histology and immunohistochemistry method for evaluation of preoperative breast core biopsies. Int J Surg pathol. 2005;13(3):247–52. doi: 10.1177/106689690501300303. [DOI] [PubMed] [Google Scholar]
- [21].Looi LM, Loh KC. Microwave-stimulated formaldehyde fixation of experimental renal biopsy tissues: computerised morphometric analysis of distortion artefacts. Malaysian J Pathol. 2005;27(1):23–7. [PubMed] [Google Scholar]
- [22].Boon ME, Gerritis PO, Moorlang HE, Neiuwenhuis, Kok LP. Formaldehyde fixation and microwave irradiation. Histochem J. 1988;20:6–7. doi: 10.1007/BF01002723. [DOI] [PubMed] [Google Scholar]
- [23].Kango PG, Deshmukh RS. Microwave processing: A boon for oral pathologists. J Oral Maxillofac Pathol. 2011;15(1):6–13. doi: 10.4103/0973-029X.80031. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [24].Panja P, Sriram G, Saraswathi TR, Sivapathasundharam B. Comparison of three different methods of tissue processing. J Oral Maxillofac Pathol. 2007;11:15–7. [Google Scholar]
- [25].Kennedy A, Foulis AK. Use of microwave oven improves morphology and staining of cryostat sections. J Clin Pathol. 1989;42:101–5. doi: 10.1136/jcp.42.1.101. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [26].Reed W, Erichsen A, Roald B. Rapid supplementary fixation in frozen sections: Microwave versus conventional fixation: A double blind comparative study. Pathology Research and Practice. 1991;187(7):824–7. doi: 10.1016/S0344-0338(11)80578-9. [DOI] [PubMed] [Google Scholar]
- [27].Hopwood D, Coghill G, Ramsay J, Milne G, Kerr M. Microwave fixation: its potential for routine techniques, histochemistry, immunocytochemistry and electron microscopy. Histochem J. 1998;16(11):1171–91. doi: 10.1007/BF01003442. [DOI] [PubMed] [Google Scholar]
- [28].Leong ASY, Milios J, Duncis CG. Antigen preservation in microwave-irradiated tissues: A comparison with formaldehyde fixation. J Pathol. 1988;156:275–82. doi: 10.1002/path.1711560402. [DOI] [PubMed] [Google Scholar]
- [29].Moune LFM, Neng LK, Ching K, Lee CE, Lee JCK. Microwave fixation in diagnostic renal pathology. J Pathol. 1987;19:17–21. doi: 10.3109/00313028709065129. [DOI] [PubMed] [Google Scholar]
- [30].Moran RA, Nelson F, Jagirdar J, Paronetto F. Application of microwave irradiation to immunohistochemistry: Preservation of antigens of the extracellular matrix. J Biotech & Histochem. 1988;63:263–9. doi: 10.3109/10520298809107599. [DOI] [PubMed] [Google Scholar]
- [31].Barrett C, Brett F, Grehan D, McDermott MB. Heat accelerated fixation and rapid dissection of the pediatric brain at autopsy: a pragmatic approach to the difficulties of organ retention. Pediatr Dev Pathol. 2004;7:595–600. doi: 10.1007/s10024-004-0041-0. [DOI] [PubMed] [Google Scholar]
- [32].Emerson LL, Tripp SR, Baird BC, Layfield LJ, L Ralph Rohr. A comparison of immunohistochemical stain quality in conventional and rapid microwave processed tissues. American Journal of Clinical Pathology. 2006;125(2):176–83. doi: 10.1309/GN6Q-CBML-LEAT-LK2M. [DOI] [PubMed] [Google Scholar]
- [33].Suri V, Chaturvedi S, Pant I, Dua R, Dua S. Application of domestic microwave for urgent histopathology reporting: an evaluation. Indian J Pathol Microbiol. 2006;49(3):348–51. [PubMed] [Google Scholar]
- [34].Steri V, Farnedi A, Montinari E, Foschini MP. Fast Track Biopsy method: a rapid approach to preoperative diagnoses. Pathologica. 2010;102:41–5. [PubMed] [Google Scholar]
- [35].Leong ASY, Duncis CG. A method of rapid fixation of large biopsy specimens using microwave irradiation. Pathology. 1986;18:222–5. doi: 10.3109/00313028609059463. [DOI] [PubMed] [Google Scholar]
- [36].Leong ASY, Daymon ME, Milios J. Microwave irradiation as a form of fixation for light and electron microscopy. J Pathol. 1985;146:313–21. doi: 10.1002/path.1711460404. [DOI] [PubMed] [Google Scholar]
- [37].Leong ASY, Price D. Incorporation of microwave tissue processing into a routine pathology laboratory: impact on turnaround times and laboratory work patterns. J Pathol. 2004;36(4):321–4. doi: 10.1080/00313020410001721564. [DOI] [PubMed] [Google Scholar]