Skip to main content
. 2013 Autumn;12(4):645–657.

Table 2.

Performance characteristics of proposed electrode at different temperatures

Temperature ( ° C) Slope (mV/decade) Usable concentration range (pAT) E ° (mV)
25 30.1 5-2 -32.5
30 30.1 5-2 -32.8
35 30 5-2 -32
40 30 5-2 -32.8
45 29.8 4.8-2 -34.4
50 29.5 4.8-2 -35.5
60 29 4.5-2 -37