Skip to main content
. 2013 Aug 16;42(8):20120445. doi: 10.1259/dmfr.20120445

Table 3.

Geometric distortion: average distance between gaps

CT apparatus Exposure conditions (kV, mA, s) Voxel size (μm3) Distances average (mm) Min–max
CS9300 60, 2, 20 90 × 90 × 90 10.05 9.90–10.30
84, 5, 20 10.04 10.00–10.10
90, 10, 20 10.04 10.00–10.10
84, 5, 12 10.19 10.10–10.20
3D Accuitomo 80 (Morita) 84, 5, 17.5 80 × 80 × 640 10.08 10.00–10.10
80 × 80 × 80 10.09 10.00–10.20
Veraviewepocs 3Df (Morita) 84, 5, 9.4 125 × 125 × 1000 10.14 10.00–10.30
125 × 125 × 125 10.13 10.00–10.30

3D, three-dimensional; max, maximum; min, minimum.