Observed minimum diameter of contrast phantom

Device Protocol Observed minimum diameter (mm) of each rod material
LDPE Delrin PTFE
CS9300 (Carestream, Rochester, NY) CS_p_0.09_60_2 × × 2
CS_p_0.09_84_5 1 2 1
CS_p_0.09_90_5 1 2 1
CS_p_0.09_84_5_fast 1 2 1
3D Accuitomo 80 (Morita, Kyoto, Japan) Accuit_p_0.64_90_5_wPMMA 1 1 1
Accuit_p_0.64_90_1_default 1 2 1
Accuit_p_0.64_90_3,5,8_default 1 1 1
Accuit_p_0.08_90_5_wPMMA 1 1 1
Accuit_p_0.08_90_1_default 1 ×  1
Accuit_p_0.08_90_3,5_default 1 2 1
Accuit_p_0.08_90_8_default 1 1 1
Veraviewepocs 3Df (Morita) Vera_p_1.0_90_5_wPMMA 1 2 1
Vera_p_1.0_90_1,3,5,8_default 1 2 1
Vera_p_0.125_90_5_wPMMA 1 2 1
Vera_p_0.125_90_1_default 2 × 1
Vera_p_0.125_90_3_default 1 3 1
Vera_p_0.125_90_5,8_default 1 2 1

LPDE, low-density polyethylene; PMMA, polymethyl methacrylate; PTFE, polytetrafluoroethylene.

“default” is when the inserts are in the default setting according to the manual. “wPMMA” is when other holes were occupied by uniform PMMA inserts. Other abbreviations are the same as in Table 2, except that the tube current is combined when the results were the same. × indicates that all rods (1–5 mm) cannot be observed.