Table 5.
Device | Protocol | Voxel size (μm3) | Maximum distinguishable line pair (LP mm−1) | |
xy plain | z plain | |||
CS9300 (Carestream, Rochester, NY) | CS_p_0.09_60_2 | 90 × 90 × 90 | 1.7 | 1 |
CS_p_0.09_84_5 | 2.5 | 2 | ||
CS_p_0.09_90_5 | 2.8 | 2 | ||
CS_p_0.09_84_5_fast | 2.5 | 2 | ||
3D Accuitomo 80 (Morita, Tokyo, Japan) | Accuit_p_0.64_90_5_wPMMA | 80 × 80 × 640 | 2 | 0 |
Accuit_p_0.64_90_5_default | 2 | 0 | ||
Accuit_p_0.08_90_5_wPMMA | 80 × 80 × 80 | 2 | 2.5 | |
Accuit_p_0.08_90_5_default | 2.5 | 2 | ||
Veraviewepocs 3Df (Morita) | Vera_p_1.0_90_5_wPMMA | 125 × 125 × 1000 | 1.7 | 0 |
Vera_p_1.0_90_5_default | 1.7 | 0 | ||
Vera_p_0.125_90_5_wPMMA | 125 × 125 × 125 | 1.7 | 2 | |
Vera_p_0.125_90_5_default | 1 | 1.7 |
PMMA, polymethyl methacrylate.
“default” is when the inserts were in the default setting according to the manual. “wPMMA” is when the other holes were occupied by uniform PMMA inserts. Other abbreviations are the same as in Table 2.