Table 1.

Gradient used and the LMWH quantity loaded for the LC-MS profiling of the LMWHs and their RO-derivatives

Enoxaparin and RO-enoxaparin Tinzaparin and RO-tinzaparin Dalteparin and RO-dalteparin

Multistep gradient
Time (min) % Eluent B Time (min) % Eluent B Time (min) % Eluent B

0 10
0 0 0 5 4 10
5 0 5 5 15 18
130 35 125 35 30 18
140 70 130 70 115 35
145 70 135 70 133 70
148 0 138 5 138 70
170 0 160 5 140 10
160 10

Sample quantity loaded, μg

15 15 30 30 30 30

Corresponding Figure

Fig. 4 and ESM Fig. S5 ESM Fig. S6 ESM Fig. S7

LC conditions: column C18 100 x 2.1 mm, 2.6 μm; eluents A and B – 10 mM DBA and 10 mM CH3COOH in H2O-CH3CN = 9:1 (v/v) and CH3CN, respectively; flow rate – 0.25 ml/min