Figure 1. Incubation setup for Experiment 1.

Figure 1

The incubation bottle was placed in a temperature controlled water bath. The 2 optodes (full range and trace) were placed on opposite sides in the upper half of the bottle. The STOX sensor was inserted through a glass tube in the center of the bottle. For calibration, known volumes of air saturated water were added through the pressure compensation port. A glass coated magnet was rotated by a magnetic stirrer below the container.