A new device to relieve venipuncture pain can affect haematology test results

Gabriel Lima-Oliveira 1,2,3,, Giuseppe Lippi 4, Gian Luca Salvagno 1, Marise Danielle Raulino Campelo 3, Katharyne Soares Adala Tajra 3, Flavio dos Santos Gomes 3, Carlos David Valentim 3, Sylvio José Colonna Romano 3, Geraldo Picheth 2, Gian Cesare Guidi 1,2
PMCID: PMC3934256  PMID: 24120583

Abstract

Background

In vitro diagnostic tests play a key role in patients’ management (e.g., guiding red blood cell transfusions). The aim of this study was to evaluate the impact of an innovative device (Buzzy®) which is claimed to be able to relieve venipuncture pain by means of cold and vibration. This device was applied during collection of venous blood by venipuncture for conventional haematology testing.

Materials and methods

Blood was drawn from 100 volunteers by a single, expert phlebotomist. A vein was located in the left forearm without applying a tourniquet but using a subcutaneous tissue transilluminator device, so that venous stasis was avoided. Blood samples were collected with a 20G straight needle directly into 4mL K3EDTA vacuum tubes. In sequence, external cold and vibration was established by Buzzy® on the right forearm −5 cm above the venipuncture site- for 1 minute before venipuncture and continued until the end of the same procedure already performed in the left forearm. Conventional haematological tests were performed using the same instrument (Sysmex® XE-2100D) in all cases.

Results

When Buzzy® was applied before drawing blood, erythrocyte counts and associated parameters (i.e., haemoglobin and haematocrit) were higher, whereas platelet number, leucocyte count and differential were lower. Statistically and clinically significant differences (P <0.001) were observed for erythrocytes, haemoglobin and haematocrit.

Discussion

From a practical perspective, cold-induced haemoconcentration promotes the efflux of water, diffusible ions and low molecular weight molecules from the vessel, thus increasing the concentration of other blood analytes at the puncture site. These variations may influence test results, especially for erythrocytes, haemoglobin and haematocrit. The novel Buzzy® device should, therefore, be used with caution when collecting blood for conventional haematological testing because of the observed bias introduced in some parameters.

Keywords: blood specimen collection, laboratory error, phlebotomy, preanalytical variability, reproducibility of results

Introduction

In vitro diagnostic testing plays a crucial role in patients’ management (e.g. for deciding whether red blood cell transfusion is necessary), provided that quality is ensured through the various phases of the testing process1,2. In agreement with the “brain-to-brain turnaround time loop” model3, which was described more than four decades ago, laboratory testing entails nine crucial phases, i.e., ordering, identification, collection, transport, preparation, analysis, test reporting, interpretation and action4. These phases have also been grouped into five higher classes: (i) pre-pre-analytical, (ii) pre-analytical, (iii) analytical, (iv) post-analytical and (v) post-post-analytical phases5. There is now clear evidence that more than two-thirds of all errors arising from the testing process occur during the pre-pre- and pre-analytical phases69. The best approach for overcoming these problems is based on clinical governance and risk-management, which represent a set of complex actions put into act to guarantee the safety of patients and healthcare staff, and which contribute to defining the quality management system10. Unfortunately, the best of our knowledge there are so far only a few routine procedures for detecting non-conformities in the pre-pre- and pre-analytical phases11,12. In the pre-analytical phase, in particular, the processes involving blood drawing have been ignored by most studies that have addressed the leading sources of errors1315.

An innovative device called Buzzy® (MMJ Labs, Atlanta, GA, USA), which combines a cooling ice-pack and a vibrating motor, is currently marketed with the claim that it may be effective at attenuating venipuncture pain in adults and children, thus ultimately enhancing patients’ compliance during venous blood collection1618. The aim of this study was to assess the impact of using Buzzy® during blood sample collection on conventional haematology tests, since the combination of cold and vibration close to the venipuncture site may be a new and potential source of pre-analytical variability.

Materials and methods

Study design

A group of 100 healthy adults of both genders, who were volunteers for this study, were enrolled from a private laboratory (Bioanalise) in the city of Teresinay, Piaui, Brazil. Each volunteer provide informed consent to inclusion in this trial, which was performed in conformity with the Declaration of Helsinki and under the terms of relevant local legislation.

Collection of diagnostic blood specimens

The blood specimens was collected in only one working day by a single, expert phlebotomist, according to established phlebotomy guidelines14,19. All subjects respected a 12-hour fast and remained seated for at least 15 minutes prior to phlebotomy, so that possible interferences of blood distribution due to the posture could be prevented20. A suitable vein was identified in the left forearm using a subcutaneous tissue transilluminator device (Venoscópio IV plus, Duan do Brasil, Sao Paulo, Brazil) without the application of a tourniquet2123. Blood samples were collected with 20G straight needles into 3 mL Venosafe® K2EDTA vacuum tubes (5.9 mg K2EDTA, ref. VF-053SDK06; Terumo, Leuven, Belgium). Nearly 2 mL of blood were preliminarily collected into a discard tube without additive (Vacuette®, Greiner Bio-One GmbH, Kremsmünster, Austria) to exclude potential interference from contact phase activation of blood, or tissue factor. In sequence, external cold and vibration was applied by Buzzy® on the right forearm, 5 cm above the puncture site, for 1 minute before venipuncture and maintained until the end of the same procedure performed in the left forearm. Buzzy® is a reusable 8×5×2.5 cm plastic device that resembles a bee and contains a battery-powered vibrating motor with an 18 g solid frozen ice-pack underneath. All blood collection steps were appropriately standardised (e.g., use of needles and vacuum tubes of the same lot).

Laboratory tests

All samples were processed in triplicate for conventional haematological testing immediately after collection (i.e., within 15 minutes). All specimens were analysed in the same Sysmex® XE-2100D, automated haematology analyser (Sysmex Corporation®, Kobe, Japan). The parameters tested included red blood cell (RBC) count, haematocrit (Hct), haemoglobin (Hb) concentration, mean corpuscular volume, mean corpuscular haemoglobin content, RBC distribution width, white blood cell (WBC) count and differential, including lymphocytes, monocytes, neutrophils, eosinophils and basophils, platelet count and mean platelet volume. The instrument was previously calibrated against appropriate proprietary reference standard material and verified with the use of appropriate controls. Our internal quality control evaluation of the within-run precision of the Sysmex® XE-2100D showed low coefficients of variation (Table I).

Table I.

Impact of Buzzy® on routine haematology tests.

Tests Units Desirable bias (%) CVa Buzzy® Gold standard Mean % difference P-value
RBC** (1012/L) 1.7 1.5 4.90±0.55 4.80±0.55 2.0 0.0006
Hb** (g/L) 1.8 1.0 141.4±13.2 137.9±12.7 2.5 0.0002
Hct** (%) 1.7 1.5 41.5±4.0 40.6±4.0 2.2 0.0005
MCV* (fL) 1.2 1.0 84.6 (81.9–88.1) 84.4 (81.8–88.3) 0.2 0.6245
RDW** (%) 1.7 2.0 12.7±0.6 12.7±0.5 0.0 0.3343
WBC** (109/L) 5.6 3.0 7.10±1.89 7.35±1.94 −3.5 0.0099
NEU** (106/L) 9.1 8.0 4.15±1.49 4.27±1.57 −2.9 0.0256
LYMP** (106/L) 7.4 8.0 2.32±0.80 2.41±0.80 −3.9 0.0425
MONO** (106/L) 13.2 9.8 0.28±0.05 0.29±0.08 −3.6 0.4431
EOS** (106/L) 19.8 15.5 0.16±0.07 0.16±0.08 0.0 0.6165
BASO** (106/L) 15.4 9.7 0.041±0.02 0.046±0.02 −12.2 0.0894
PLT** (109/L) 5.9 4.0 272±66 274±66 −0.7 0.3705
MPV** (fL) 2.3 1.2 9.09±0.71 9.12±0.81 −s0.3 0.7116
*

Non-normal distribution; the values are presented as median (interquartile range); P-value represents the significance by Wilcoxon ranked-pairs test.

**

Normal distribution; the values are presented as mean ± standard deviation; P-value represents the significance by paired Student’s t-test. The P-values in bold are statistically significant (P <0.001) and bold mean % differences represent clinically significant variations, when compared with desirable bias24.

RBC: red blood cell count; Hct: haematocrit; Hb: haemoglobin concentration, MCV: mean corpuscular volume; RDW: red blood cell distribution width; WBC: white blood cell count; NEU: neutrophil count; LYMP: lymphocyte count; MONO: monocyte count; EOS: eosinophil count; BASO: basophil count; PLT: platelet count; MPV: mean platelet volume.

Statistical analysis

The significance of differences between samples was assessed by using a Student’s paired t-test after the normal distribution of values had been verified with the D’Agostino-Pearson omnibus test. A non-normal distribution was found only for mean corpuscular volume: in this case, results were assessed by the Wilcoxon’s ranked-pairs test. The level of statistical significance was set at P <0.001. The bias from conventional and Buzzy-assisted venipuncture was assessed according to the current desirable quality specifications for bias (B), derived from biological variation24.

Results

When cold and vibration were established by Buzzy® before collection of blood samples, RBC count and associated parameters (Hb and Hct) showed significantly increased values, whereas the platelet count, WBC count and differential counts were decreased (Table I). The differences were statistically (P <0.001) and clinically significant only for RBC count, Hb concentration and Hct.

Discussion

It has recently been shown that direct-to-consumer testing presents more risks than real opportunities for patients and healthcare systems4. Our study group also previously demonstrated that changes of brand of in vitro diagnostic devices (e.g. vacuum tubes2528 and syringes29) represent a new source of pre-analytical variability. It is also noteworthy that the most important guideline on diagnostic blood specimen collection by venipuncture recommends procedures that may generate venous stasis14. Inal and Kelleci, and Baxter et al. recently showed that Buzzy® is effective at relieving needle pain during venipuncture1618. Obviously, the ability of phlebotomists to put patients at ease is considered a very important issue by outpatients30. Oatey and Stiller showed that 28% of inpatients shared the opinion that venipuncture is indeed a painful medical procedure31. As such, the Buzzy® device may be viewed as an attractive novelty to overcome this important clinical issue. Indeed, as suggested by the manufacturer of Buzzy®32 the phlebotomist should distract the patients for a while (15 sec) after placing the device. We measured that the time lap needed to perform this pre-phlebotomy operation was 55±10 sec (mean ± SD for 8 plebotomists); we, therefore, adopted 1 minute as our standardised time. A previous assessment of Buzzy® use during diagnostic specimen collection by venipuncture for routine immunochemistry testing showed negligible variations, with significant bias only reported for total protein and albumin concentrations out of the 34 analytes tested33. Nevertheless, it is important to focus on RBC count, Hb concentration and Hct (Table I), since these parameters are often and misleadingly considered very solid and relatively insensitive to pre-analytical variables. Obviously the quality specifications for bias derived from biological variation (Table I)24 are considered both very important and useful in daily practice by quality managers of medical laboratories3437. According to our results, there is a tangible risk that some physicians could make inappropriate clinical decisions (e.g., delay or avoid RBC transfusions3840) as a consequence of biased24 haematological results due to the use of the new Buzzy® device. The most reasonable explanation for the bias is vasoconstriction due to direct skin cooling41,42. From a practical perspective, the cold-induced haemoconcentration43 promotes efflux of water, diffusible ions and low molecular weight molecules from the vessel and may, therefore, increase the concentration of various analytes in close proximity to the venipuncture site. These variations could potentially impair the reliability of test results, especially RBC count, Hb concentration and Hct. In a systematic review Fuller et al. showed that vibration increases muscle perfusion, with the magnitude of the increase being positively related to the vibratory load applied44. The finding of a decreased WBC count is consistent with the hypothesis that the external vibration may trigger leucocyte recruitment.

It is well established that the sampling process is a crucial step in research45. One strength of our study was the accurate standardisation of all steps (i.e., all diagnostic blood specimens were collected by a single expert phlebotomy in one working day and all in vitro equipment was from the same lot), which minimised potential biases.

In conclusion the novel Buzzy® device should be used with caution when collecting blood for conventional haematological testing because of the observed bias introduced in some parameters. Another important issue that may argue against the use of this device is the risk of triggering precipitation of cold agglutinins as a consequence of prolonged contact of ice with the skin. This aspect deserves scrutiny in further investigations.

Acknowledgement

The Authors acknowledge all clinical biochemistry staff from Clinical Laboratory Bioanalise, Piaui (Brazil), for their skillful technical support. Special thanks for Dr. Risemberg Soares Pereira, Dr. Naiza Silva Ribeiro and M.T. Lícia Maria Rodrigues Lustosa.

Footnotes

Disclosure

All authors confirm that they contributed to the intellectual content of this paper and met the following three requirements: (i) significant contributions to the conception and design, acquisition of data, or analysis and interpretation of data; (ii) drafting or revising the article for intellectual content; and (iii) final approval of the published article.

The Authors do not have any potential conflicts of interest relevant to this article.

References

  • 1.Plebani M, Lippi G. To err is human. To misdiagnose might be deadly. Clin Biochem. 2010;43:1–3. doi: 10.1016/j.clinbiochem.2009.07.002. [DOI] [PubMed] [Google Scholar]
  • 2.Hallworth M, Hyde K, Cumming A, Peake I. The future for clinical scientists in laboratory medicine. Clin Lab Haematol. 2002;24:197–204. doi: 10.1046/j.1365-2257.2002.00455.x. [DOI] [PubMed] [Google Scholar]
  • 3.Lundberg G. Acting on significant laboratory results. J Am Med Assoc. 1981;245:1762–3. doi: 10.1001/jama.1981.03310420052033. [DOI] [PubMed] [Google Scholar]
  • 4.Lippi G, Favaloro EJ, Plebani M. Direct-to-consumer testing: more risks than opportunities. Int J Clin Pract. 2011;65:1221–9. doi: 10.1111/j.1742-1241.2011.02774.x. [DOI] [PubMed] [Google Scholar]
  • 5.Plebani M, Lippi G. Closing the brain-to-brain loop in laboratory testing. Clin Chem Lab Med. 2011;49:1131–3. doi: 10.1515/CCLM.2011.617. [DOI] [PubMed] [Google Scholar]
  • 6.Wallin O, Soderberg J, Van Guelpen B, et al. Preanalytical venous blood sampling practices demand improvement--a survey of test-request management, test-tube labelling and information search procedures. Clin Chim Acta. 2008;391:91–7. doi: 10.1016/j.cca.2008.02.016. [DOI] [PubMed] [Google Scholar]
  • 7.Lippi G, Bassi A, Brocco G, et al. Preanalytic error tracking in a laboratory medicine department: results of a 1-year experience. Clin Chem. 2006;52:1442–3. doi: 10.1373/clinchem.2006.069534. [DOI] [PubMed] [Google Scholar]
  • 8.Carraro P, Plebani M. Errors in a stat laboratory: types and frequencies 10 years later. Clin Chem. 2007;53:1338–42. doi: 10.1373/clinchem.2007.088344. [DOI] [PubMed] [Google Scholar]
  • 9.Plebani M, Carraro P. Mistakes in a stat laboratory: types and frequency. Clin Chem. 1997;43:1348–51. [PubMed] [Google Scholar]
  • 10.Bambi F, Spitaleri I, Verdolini G, et al. Analysis and management of the risks related to the collection, processing, and distribution of peripheral blood haematopoietic stem cells. Blood Transfus. 2009;7:3–17. doi: 10.2450/2008.0006-08. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Lippi G, Fostini R, Guidi GC. Quality improvement in laboratory medicine: extra-analytical issues. Clin Lab Med. 2008;28:285–94. doi: 10.1016/j.cll.2007.12.007. [DOI] [PubMed] [Google Scholar]
  • 12.Lippi G, Guidi GC. Risk management in the preanalytical phase of laboratory testing. Clin Chem Lab Med. 2007;45:720–7. doi: 10.1515/CCLM.2007.167. [DOI] [PubMed] [Google Scholar]
  • 13.Lima-Oliveira G, Guidi GC, Salvagno GL, et al. Is phlebotomy part of the dark side in the clinical laboratory struggle for quality? Lab Med. 2012;43:17–21. [Google Scholar]
  • 14.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Impact of the phlebotomy training based on CLSI/NCCLS H03-A6 - procedures for the collection of diagnostic blood specimens by venipuncture. Biochem Med (Zagreb) 2012;22:342–51. doi: 10.11613/bm.2012.036. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Lippi G, Guidi GC. Preanalytic indicators of laboratory performances and quality improvement of laboratory testing. Clin Lab. 2006;52:457–62. [PubMed] [Google Scholar]
  • 16.Inal S, Kelleci M. Relief of pain during blood specimen collection in pediatric patients. MNC Am J Matern Child Nurs. 2012;37:339–45. doi: 10.1097/NMC.0b013e31825a8aa5. [DOI] [PubMed] [Google Scholar]
  • 17.Baxter AL, Cohen LL, McElvery HL, et al. An integration of vibration and cold relieves venipuncture pain in a pediatric emergency department. Pediatr Emerg Care. 2011;27:1151–6. doi: 10.1097/PEC.0b013e318237ace4. [DOI] [PubMed] [Google Scholar]
  • 18.Baxter AL, Leong T, Mathew B. External thermomechanical stimulation versus vapocoolant for adult venipuncture pain: pilot data on a novel device. Clin J Pain. 2009;25:705–10. doi: 10.1097/AJP.0b013e3181af1236. [DOI] [PubMed] [Google Scholar]
  • 19.Clinical Laboratory Standards Institute. CLSI H3-A6 document. 6th ed. Wayne, PA: Clinical Laboratory Standards Institute; 2007. Procedures for the collection of diagnostic blood specimens by venipuncture. [Google Scholar]
  • 20.Guder WG, Narayanan S, Wisser H, Zawta B. Diagnostic samples: from the patient to the laboratory: the impact of preanalytical variables on the quality of laboratory results. 4th ed. Wiley-Blackwell; 2009. [Google Scholar]
  • 21.Lima-Oliveira G, Lippi G, Salvagno GL, et al. New ways to deal with known preanalytical issues: use of transilluminator instead of tourniquet for easing vein access and eliminating stasis on clinical biochemistry. Biochem Med (Zagreb) 2011;21:152–9. doi: 10.11613/bm.2011.024. [DOI] [PubMed] [Google Scholar]
  • 22.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Transillumination: a new tool to eliminate the impact of venous stasis during the procedure for the collection of diagnostic blood specimens for routine haematological testing. Int J Lab Hematol. 2011;33:457–62. doi: 10.1111/j.1751-553X.2011.01305.x. [DOI] [PubMed] [Google Scholar]
  • 23.Lima-Oliveira G, Salvagno GL, Lippi G, et al. Elimination of the venous stasis error for routine coagulation testing by transillumination. Clin Chim Acta. 2011;412:1482–4. doi: 10.1016/j.cca.2011.04.008. [DOI] [PubMed] [Google Scholar]
  • 24.Ricos C, Alvarez V, Cava F, et al. Current databases on biological variation: pros, cons and progress. Scand J Clin Lab Invest. 1999;59:491–500. doi: 10.1080/00365519950185229. [DOI] [PubMed] [Google Scholar]
  • 25.Lima-Oliveira G, Lippi G, Salvagno GL, et al. K3EDTA vacuum tubes validation for routine hematological testing. ISRN Hematology. 2012;2012:875357. doi: 10.5402/2012/875357. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Pre analytical management: serum vacuum tubes validation for routine clinical chemistry. Biochem Med (Zagreb) 2012;22:180–6. doi: 10.11613/bm.2012.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Sodium citrate vacuum tubes validation: preventing preanalytical variability in routine coagulation testing. Blood Coagul Fibrinolysis. 2013;24:252–5. doi: 10.1097/MBC.0b013e32835b72ea. [DOI] [PubMed] [Google Scholar]
  • 28.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Brand of dipotassium EDTA vacuum tube as a new source of preanalytical variability in routine haematology testing. Br J Biomed Sci. 2013;70:6–9. doi: 10.1080/09674845.2013.11669922. [DOI] [PubMed] [Google Scholar]
  • 29.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Different manufacturers of syringes: a new source of variability in blood gas, acid-base balance and related laboratory test? Clin Biochem. 2012;45:683–7. doi: 10.1016/j.clinbiochem.2012.03.007. [DOI] [PubMed] [Google Scholar]
  • 30.Cembrowski GS, Strauss S, Waldeland LJ, et al. Are phlebotomy services completely satisfying our patient customers? 1995 Institute: Frontiers in Laboratory Practice Research. 1996:198–208. [Google Scholar]
  • 31.Oatey A, Stiller K. An evaluation of the level of satisfaction with a dedicated inpatient service at an rehabilitation centre. Int J Nurs Pract. 2009;15:553–9. doi: 10.1111/j.1440-172X.2009.01779.x. [DOI] [PubMed] [Google Scholar]
  • 32.MMJ Labs. Buzzy Personal Pain Contol: How to use Buzzy. [Accessed on 05/11/2012]. Available at: http://www.buzzy4shots.com/Resources/how-to-use-buzzy.html.
  • 33.Lima-Oliveira G, Lippi G, Salvagno GL, et al. Quality impact on diagnostic blood specimen collection using a new device to relieve venipuncture pain. Indian J Clin Biochem. 2013;28:235–41. doi: 10.1007/s12291-013-0319-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Ricos C, Cava F, Garcia-Lario JV, et al. The reference change value: a proposal to interpret laboratory reports in serial testing based on biological variation. Scand J Clin Lab Invest. 2004;64:175–84. doi: 10.1080/00365510410004885. [DOI] [PubMed] [Google Scholar]
  • 35.Westgard J. Biological Variation Database Specifications. 2012. [Accessed on 01/12/2012]. Available at: http://www.westgard.com/biodatabase1.htm.
  • 36.Cembrowski GS, Tran DV, Higgins TN. The use of serial patient blood gas, electrolyte and glucose results to derive biologic variation: a new tool to assess the acceptability of intensive care unit testing. Clin Chem Lab Med. 2010;48:1447–54. doi: 10.1515/CCLM.2010.286. [DOI] [PubMed] [Google Scholar]
  • 37.Plebani M, Lippi G. Biological variation and reference change values: an essential piece of the puzzle of laboratory testing. Clin Chem Lab Med. 2012;50:189–90. doi: 10.1515/cclm.2011.751. [DOI] [PubMed] [Google Scholar]
  • 38.De Leon EM, Szallasi A. “Transfusion indication RBC (PBM-02)”: gap analysis of a Joint Commission Patient Blood Management Performance Measure at community hospital. Blood Transfus. 2012 doi: 10.2450/2012.0088-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Marik P, Corwin H. Efficacy of red blood cell transfusion in the critically ill: a systematic review of the literature. Crit Care Med. 2008;36:2667–74. doi: 10.1097/CCM.0b013e3181844677. [DOI] [PubMed] [Google Scholar]
  • 40.Guzzetta NA. Benefits and risks of red blood cell transfusion in pediatric patients undergoing cardiac surgery. Paediatr Anaesth. 2011;21:504–11. doi: 10.1111/j.1460-9592.2010.03464.x. [DOI] [PubMed] [Google Scholar]
  • 41.Johnson J. Mechanisms of vasoconstriction with direct skin cooling in humans. Am J Physiol Heart Circ Physiol. 2007;292:H1690–1. doi: 10.1152/ajpheart.00048.2007. [DOI] [PubMed] [Google Scholar]
  • 42.Pergola P, Kellogg D, Johnson J, et al. Role of sympathetic nerves in the vascular effects of local temperature in human forearm skin. Am J Physiol Heart Circ Physiol. 1993;265:H785–92. doi: 10.1152/ajpheart.1993.265.3.H785. [DOI] [PubMed] [Google Scholar]
  • 43.Austin A, Patterson S, Kanel RV. Hemoconcentration and hemostasis during acute stress: interacting and independent effects. Ann Behav Med. 2011;42:153–73. doi: 10.1007/s12160-011-9274-0. [DOI] [PubMed] [Google Scholar]
  • 44.Fuller J, Thomson R, Howe P, Buckley J. Effect of vibration on muscle perfusion: a systematic review. Clin Physiol Funct Imaging. 2013;33:1–10. doi: 10.1111/j.1475-097X.2012.01161.x. [DOI] [PubMed] [Google Scholar]
  • 45.Simundic AM. Bias in research. Biochem Med (Zagreb) 2013;23:12–5. doi: 10.11613/BM.2013.003. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Blood Transfusion are provided here courtesy of SIMTI Servizi

RESOURCES