Skip to main content
. 2013 Oct 15;77(3):309–316. doi: 10.1007/s10337-013-2584-6

Table 1.

Method used in Tracer GC Ultra for two columns when TD was connected

Method description Carbograph PLOT Al2O3-PLOT-S
Initial temperature 35 °C 100 °C
Initial time 5 min 0.5 min
Ramp rate 1 20 °C min−1 30 °C min−1
Final temperature 1 50 °C 130 °C
Hold time 1 10 min 6.5 min
Ramp rate 2 2.5 °C min−1 30 °C min−1
Final temperature 2 65 °C 140 °C
Hold time 2 1 min 23 min
Ramp rate 3 20 °C min−1 20 °C min−1
Final temperature 3 200 °C 190 °C
Hold time 3 4 min 5 min
Carrier gas flow rate 1.5 mL min−1 1 mL min−1