Skip to main content
. 2013 Jun 5;2:361. doi: 10.1038/bonekey.2013.95

Table 1. Comparison of commercially available sclerostin assays.

Assay source Assay type Primary antibody Secondary antibody Intra-assay CV (%) Inter-assay CV (%) Detection limit (pmol/l) Detection range (pmol/l)
Biomedica ELISA Polyclonal Monoclonal 5 6 2.6 0–240
TECO ELISA Polyclonal Monoclonal <10 <10 5.7 11–176
MSD ECL Polyclonal Polyclonal 6 10 0.1 1–440

Abbreviations: CV, coefficient of variation; ECL, electro-chemiluminescence assay; ELISA, enzyme-linked immunosorbent assay; MSD, Meso Scale Discovery; TECO, TECO Medical Group.