. 2014 Jan-Feb;21(1):31–35. doi: 10.1155/2014/271061

TABLE 3.

Sensitivities and specificities of automated and manually scored PSS-AL AHI compared with PSG AHI

PSS-AL AHI PSG SDB (AHI ≥ 1.5 events/h)
PSG SDB (AHI ≥5 events/h)
Sensitivity Specificity PPV NPV Sensitivity Specificity PPV NPV
Manual PSS-AL (cut-off ≥ 1.5 events/h) 94 16 51 75 100 13 16 100
Manual PSS-AL (cut-off ≥ 5 events/h) 94 61 70 92 100 40 22 100
Automated PSS-AL (cut-off ≥ 1.5 events/h) 94 0 47 0 100 3 15 100
Automated PSS-AL (cut-off ≥ 5 events/h) 94 44 62 88 100 30 19 100

Data presented as %. AHI Apnea-hypopnea index; NPV Negative predictive value; PSG Polysomnography; PSS-AL Portable sleep study-ApneaLink (ResMed, USA); PPV Positive predictive value; SDB Sleep disordered breathing