Skip to main content
. Author manuscript; available in PMC: 2015 Mar 1.
Published in final edited form as: Ann Biomed Eng. 2013 Oct 26;42(3):503–514. doi: 10.1007/s10439-013-0930-3

Table 3.

Intra-aneurysmal WSS with Patient-Specific vs. Simplified Boundary Conditions

Hemodynamic Variable Patient-Specific Simplified p value
Mean WSS (Pa) 1.85 ± 1.34 4.06 ± 3.57 < .01
Peak WSS (Pa) 4.41 ± 3.41 8.39 ± 7.50 < .01
Normalized Mean WSS (-) 1.98 ± 1.63 2.70 ± 2.38 < .01
Normalized Peak WSS (-) 2.29 ± 2.02 3.02 ± 2.58 < .01