Abstract
Patellofemoral crepitus and clunk syndrome are infrequent, yet troublesome complications of total knee arthroplasty with a reported incidence of 0%-18%. They are primarily associated with implantation of posterior cruciate substituting designs. These entities are the result of peripatellar fibrosynovial hyperplasia at the junction of the superior pole of the patella and the distal quadriceps tendon which becomes entrapped within the superior aspect of the intercondylar box of the femoral component during knee flexion. When the knee extends, a crepitant sensation occurs as the fibrosynovial tissue exits the intercondylar box. Numerous etiologies have been proposed such as femoral component designs with a high intercondylar box ratio, previous knee surgery, reduced patellar tendon length, thinner patellar components, reduced patella-patellar component composite thickness, and smaller femoral components. Preventative measures include choice of femoral components with a reduced intercondylar box ratio, use of thicker patellar components, avoidance of over-resection of the patella, and debridement of the fibrosynovial tissue at the time of knee arthroplasty. Most patients with crepitus are unaware of the problem or have minimal symptoms so that no treatment is required. If significant disability is incurred, symptoms can be eliminated in a high percentage of patients with arthroscopic debridement of the fibrosynovial hyperplasia.
Keywords: Patellar crepitus, Total knee arthroplasty, Complication
The excellent long term clinical results of total knee arthroplasty (TKA) utilizing a posterior stabilized (PS) prosthesis have been well documented.1,2) However, despite these results, patellofemoral complications remain a common source of pain and dysfunction.3,4) In 1982, Insall et al.5) described a case of anterior knee pain associated with a peripatellar fibrous nodule that resolved following operative resection. In this report, two additional cases were proposed to have similar pathology. Conservative management was rendered. In 1989, patellar clunk syndrome was first described by Hozack et al.6) in a cases series of patients treated with the Insall-Burstein I (IB-I) prosthesis (Zimmer, Warsaw, IN, USA), a first generation PS component. While cases of patellar clunk syndrome have been reported with the use of both cruciate retaining (CR) and patellofemoral replacement (PF) prostheses,7,8) this complication has primarily occurred with use of PS TKAs with a reported incidence of 0%-18%.3,5,9-22)
While patella clunk syndrome and patellofemoral crepitus are separate entities, they share a common pathophysiology with widely varying clinical presentations. The symptoms range from a painless subtle crepitation to a painful, palpable and audible clunk. These entities are the result of peripatellar fibrosynovial hyperplasia at the junction of the superior pole of the patella and the distal quadriceps tendon (Fig. 1). In the case of patellar clunk, a discrete fibrosynovial nodule becomes entrapped within the intercondylar box of the PS femoral component during knee flexion. Subsequently, when the knee is extended to within 30-45 degrees of full extension, the nodule dislodges resulting in an audible and often painful clunk (Fig. 2). In the case of patellofemoral crepitus, the fibrosynovial hyperplasia does not develop into a discrete nodule and no audible clunk is present. Conversely, these patients complain of anterior knee pain and a continued grinding sensation with loading of the knee in a 30-60 degree range. In either case, symptoms typically develop within 2 months to 2 years following TKA.5,6,10-13,15,17-19,23) The natural history of patellofemoral crepitus and patellar clunk syndrome reveal that up to 50% of patients improve with conservative treatment and the majority are unaware the condition exists.3,12,23) However, in both syndromes, the symptoms can be disabling warranting surgical intervention with either an arthroscopic or open debridement.24-30) Nevertheless, prevention is preferred.
In this chapter, we review the etiological factors of patellar crepitus and clunk syndrome and the preventative measures to avoid its development.
HISTORICAL ETIOLOGIES
While the etiology of patellofemoral crepitus and clunk syndrome are multifactorial, the design and position of the femoral component are key factors in its development.5,9-14,17-22) First generation PS designs included a high transition zone from the trochlear groove to the intercondylar box with an abrupt transition to the distal femoral articular geometry (Fig. 3).10) It was theorized that this design feature led to the apposition of the distal quadriceps tendon to the anterior edge of the intercondylar box, resulting in soft tissue irritation, fibrosynovial hyperplasia and subsequent crepitus or clunk. Fukunaga et al.11) further characterized this transition zone by defining the intercondylar box ratio, as the intercondylar box height versus the anterior-posterior height of the femoral component (Fig. 4). Second and third generation PS femoral components have typically been designed with a decreased intercondylar box ratio with a resultant decrease in the incidence of patellar clunk and crepitus (Table 1). Notably, those prostheses with an intercondylar box ratio of less than 0.7 were found to have no patellar clunk.11) Femoral components with a high intercondylar box ratio allow contact of the distal quadriceps tendon earlier in flexion than components with a lower ratio. The authors concluded that this design feature may be responsible for the high incidence of patellofemoral crepitus and clunk with these prostheses.
Table 1.
PS: posterior stabilized.
The medial-lateral geometry of the intercondylar box has also been implicated as a causative factor. Pollock et al.17) evaluated the incidence of soft tissue entrapment within the intercondylar box of three different PS TKA designs (Fig. 5). The incidence was 3.8% with the AMK PS (DePuy, Warsaw, IN, USA), 13.5% for the AMK Congruency (DePuy), and 0% for the PFC Sigma PS (DePuy). The authors concluded that both the height and the width of the intercondylar box were design features leading to synovial entrapment.
Many other aspects of the femoral component design have been reported as etiologic factors for patellofemoral crepitus and clunk. Lonner et al.13) compared the incidence of patellar clunk syndrome in patients who received two different generation PS TKA designs, the Insall-Burstein II (IB-II, Zimmer) and the NexGen LPS (NG-LPS, Zimmer). The incidence was 4% in the IB-II group and 0% in the NG-LPS group. The modifications made to the NG-LPS femoral component were a raised lateral flange, deepened trochlear groove, and a lengthened trochlear groove resulting in a more posteriorly positioned intercondylar box and a reduced intercondylar box ratio. The authors reported that these design features allow the patella to be engaged in the trochlear groove for a greater arc of motion, thus reducing the potential for frictional quadriceps tendon irritation and the subsequent development of patellar crepitus or clunk. In a similar study, Maloney et al.19) found an increased incidence of patellar clunk in patients receiving the IB-II prosthesis (4%) compared to the Advance PS prosthesis (0%). The authors cited the posteriorly extended trochlear groove of the Advance PS as the differentiating design feature that led to the elimination of patellar clunk syndrome.
Aside from the femoral component design features, many authors suggest that patellofemoral crepitus and clunk occur are a result of intraoperative technical errors.4,6,9,14,16,23) The incidence of patellofemoral complications in patients with PS TKAs has long been attributed to patellar baja and anterior placement of the tibial tray.4) Recently, Yau et al.16) conducted a retrospective radiographic review of the extensor mechanism to identify possible causes of patellar clunk syndrome. In a consecutive series of 236 primary TKAs, 27 cases (11.4%) with patellar clunk syndrome were identified. The patients without patellar clunk served as a control group. Postoperative true lateral radiographs were utilized to evaluate the Insall-Salvati ratio, the relative anterior-posterior position of the tibial tray (Fig. 6), the height of the joint line, the patellar button height and the position of the proximal pole of the patella relative to the distal part of the femoral component (Fig. 7). A lower Insall-Salvati ratio, lower patellar component height, lower position of the proximal pole of the patella, and anteriorization of the tibial tray were all associated with a higher incidence of patellar clunk syndrome (Table 2). These data demonstrate that postoperative patella baja and anterior placement of the tibial tray is significantly related to the development of patellar clunk syndrome. In this study, there was no statistical difference in the elevation of the joint line between the two groups, which may suggest that the relative patellar baja is secondary to either inferior placement of the patellar component or an excessive distal femoral cut. In either case, the altered patellofemoral kinematics may be in part responsible for the development of patellar clunk syndrome.
Table 2.
Patellar crepitus and clunk syndrome have also been attributed to abnormal patellar tracking secondary to inadequate soft-tissue balance or tibial component malrotation.4) In a second part of their study, Yau et al.16) evaluated the congruency of the patellar component with the trochlear groove in patients who underwent bilateral TKAs but developed unilateral patellar clunk syndrome. They found an average patellar tilt of 9.59° ± 3.73° in patient with patellar clunk and (5.95° ± 3.04°) in patients without patellar clunk (p = 0.019). Fukunaga et al.11) similarly found an association between the incidence of patellar clunk and patellar mal-tracking. In their series, the mean patellar tilt was 6.00° ± 3.4° and 4.08° ± 2.9° in patients with and without patellar clunk respectively (p = 0.020). The accompanying logistic regression analysis demonstrated that for every 1° increase in the postoperative patellar tilt, there is an increase the incidence of patellar clunk by 1.27.
Position of the patellar component has also been implicated as a cause of patellar clunk syndrome.6) In their original case series, Hozack et al.6) described one case where proximal overhang of the patellar button beyond the superior border of the patella led to a catching of the component on the anterior portion of the intercondylar box. Trimming of the overhanging portion of the button allowed complete resolution of symptoms. Subsequent studies have not supported this etiology. However, the pathophysiology may be applicable to occurrence of patellar clunk in TKA without patellar resurfacing. Shoji and Shimozaki21) reported 11 cases of patellar clunk syndrome noted intraoperatively during PS TKA without patellar resurfacing. In this study, the increased medial to lateral width and intercondylar box ratio of the prosthesis (AGC; Biomet, Warsaw, IN, USA) allowed the superior pole of the patella to catch within the intercondylar box in a similar manner as the fibrosynovial nodule proposed by other authors. Shaving the superior pole of the patella eliminated the clunk in seven knees.
While many authors have identified component design and positioning as risk factors for the development of patellar crepitus, few have implicated the association of increased postoperative knee flexion.15,18) Schroer et al.15) retrospectively reviewed 747 consecutive TKAs in three contiguous cohorts for the occurrence of patellar clunk syndrome. In the first group, performed with the use of a standard median parapatellar arthrotomy and the Ascent PS (Biomet) femoral component, the incidence of patellar clunk was 0.5%. In the second group utilizing a mini-subvastus (MIS) approach, the same femoral component, and a resultant increase in the mean postoperative flexion, (113° vs. 124°) the incidence of patellar clunk syndrome was 6%. In the third group, they maintained the MIS surgical approach, transitioned to using a third generation femoral component (Vanguard PS, Biomet) with a resultant decrease in the incidence of patellar clunk to 0.4% despite the fact higher levels of flexion were obtained (125°). The preponderance of patellar clunk in the second group was attributed to the combination of increased mean postoperative flexion while using a second generation femoral component. These authors hypothesize that increased knee flexion positions the distal quadriceps tendon further distally and posteriorly on the femoral component, thus increasing the likelihood of soft tissue impingement in the intercondylar box. Third generation femoral components have a more posteriorly positioned intercondylar box and a more gradual transition to the distal femoral geometry, thus decreasing soft tissue impingement despite allowing higher levels of postoperative knee flexion. Similarly, patients with stiffness following TKA with a second generation femoral component rarely develop patellar crepitus since their flexion is not sufficient to allow the fibrosynovial tissue to become entrapped in the intercondylar space.18)
CONTROL-MATCHED EVALUATION OF PAINFUL PATELLAR CREPITUS
Since its original description by Hozack et al.6) in 1989, numerous studies have added to the long list of causes of patellar crepitus and clunk. However, this multifactorial nature makes it challenging for surgeons to identify with confidence areas of intervention to prevent its development as well as identify preoperatively patients who may be at high risk. Dennis et al.23) conducted a large retrospective, multicenter case-control study to identify the onset of patellar crepitus, as well as clinical, radiographic, and surgical variables that increase the risk of patients developing painful, symptomatic patellar crepitus following PS TKA (Table 3).
Table 3.
*Patient clinical and radiographic variables were treated as continuous variables. †Surgical variables were treated as categorical variables.
The surgical databases of two institutions were searched to identify all patients who required surgical intervention for symptomatic patellar crepitus or clunk after PS TKA. Between 2002 and 2008 over 4,000 PS TKAs from the same implant system (PFC Sigma PS TKA [non-HP design]; Depuy) were identified, of which 60 met inclusion criteria. Sixteen (26.7%) had patellar clunk, whereas the remaining 44 (73.3%) developed patellar crepitus. Study patients were matched for age (± 3 years), gender, and body mass index (± 3 points) with a control patient group with well-functioning TKAs who did not experience patellar crepitus or clunk. A univariate analysis of the clinical, radiographic, and surgical parameters (Table 3) was completed to identify variables for a multivariate logistic regression model. Additionally, variables previously reported in the literature with clinical or surgical plausibility were also included in the regression model.
The average time of onset of the symptoms in the patellar crepitus group was 10.9 months (4 to 27 months). The univariate analysis identified 6 factors to statistically increase the risk of patellar clunk or crepitus: a reduced preoperative (54.54 mm vs. 57.90 mm; p = 0.0089) or postoperative (54.56 mm vs. 58.10 mm; p = 0.0089) patellar tendon length, a thinner postoperative composite patellar component thickness (22.33 mm vs. 23.24 mm; p = 0.0219), a reduced patellar component size (p = 0.011), an increased posterior femoral condylar offset (1.27 mm versus 0.17 mm; p = 0.023), and a history of pervious knee surgery (p = 0.011) (Table 4). The best fit multivariate model showed increased risk of developing patellar crepitus in patients with previous knee surgery (p = 0.0009), smaller femoral component size (size < 3; p = 0.0253), thicker polyethylene bearings (> 12.5 mm; p = 0.0015), reduced preoperative patellar tendon length (p = 0.0098), increased femoral component flexion (p = 0.0282), and an increase in the postoperative posterior femoral condylar offset (p < 0.001) (Table 5).
Table 4.
Table 5.
The results of this study further support many of the historically reported etiologies and risk factors for patellar crepitus and clunk. Numerous studies have implicated first and second generation femoral component designs in patellar crepitus,5,9-22) especially those with an intercondylar box ratio of greater than 0.7,11) like the PFC Sigma PS TKA (0.85-0.87) utilized in this study.
A decreased postoperative patellar tendon length and Insall-Salvati ratio have been cited by other authors as an etiologies for patellar crepitus.4,16) The above data from the univariate analysis confirm that a decreased patellar height and shorter patellar tendon length both preoperatively and postoperatively are associated with the development of patellar crepitus. Moreover, the multivariate analysis demonstrated that a relative shortening of the patellar tendon via the use of thicker polyethylene inserts was significant risk factor. The authors suggest that these conditions increase the contact forces between the distal quadriceps tendon and the intercondylar box resulting in tendon irritation, development of fibrosynovial tissue proliferation, and eventual patellar crepitus or clunk. A reduced composite patella-patellar component thickness and use of a smaller (thinner) patellar component decrease the offset of the distal quadriceps tendon from the superior aspect of the intercondylar box, thereby increasing the risk of frictional irritation of the posterior aspect of the distal quadriceps tendon, fibrosynovial tissue development, and subsequent patellar crepitus.
Figgie et al.4) found a correlation of anterior placement of the femoral component and the use of an undersized femoral component with patellofemoral complications following PS TKA. The above data do not support anterior placement of the femoral component as a risk factor. However, the multivariate analysis demonstrated a strong correlation between both the use of a small femoral component as well as a flexed posture of the femoral component with the development of patellar crepitus. They suggest the latter may irritate the quadriceps tendon as it traverses the proximal aspect of the anterior flange of the femoral component.
No prior study has made the correlation of increased posterior femoral condylar offset and the development of patellar crepitus. The authors speculate that increasing the posterior condylar offset may result in a relative anterior shift of the intercondylar box, thus predisposing to quadriceps tendon irritation. The authors speculate the increased risk of patellar crepitus in subjects with prior knee surgery is secondary to development of increased intra-articular fibrosis from earlier knee operative procedures. This arthrofibrosis likely reduces soft tissue compliance, possibly reducing patellar height and increasing tension of quadriceps tendon as it traverses the intercondylar box of the femoral component.
The authors emphasize that this data cannot dispute other causes of postoperative patellar crepitus previously reported in the literature that were not found to be statistically significant given that the patient volume may not be adequately powered for each variable. Nevertheless, their data both supports many of the previous investigations as well as adding many new potential etiologies and risk factors for the development of patellar crepitus. In summary, the central pathophysiological hypothesis is that these factors in a multifactorial fashion result in increased contact forces and subsequent irritation of the distal quadriceps tendon as it traverses the superior aspect of the intercondylar box of a PS femoral component.
CONTROL-MATCHED COMPUTATIONAL EVALUATION OF TENDO-FEMORAL CONTACT
To further evaluate the risk factors reported in the above discussed analysis, Hoops et al.31) performed a control matched computational evaluation of tendo-femoral contact in patients implanted with a PS TKA. Their objective was to determine the influence of variations in patellar tendon length, flexion-extension alignment of the femoral component, position of the joint line, and patellar button size on the tendo-femoral contact and tendon-to-intercondylar notch spacing during a simulated deep knee bend activity.
The computational model was adapted from an in vitro experiment where a cadaveric knee was implanted with a PS TKA (PFC Sigma; DePuy) and subsequently mounted on the Kansas Knee Simulator (KKS). The KKS then facilitated a simulated deep knee bend activity. Magnetic resonance imaging of the implanted knee allowed for digitization of the implant surfaces and composite three-dimensional modeling. The clinical parameters were then varied by two standard deviations in both directions from the neutral implanted alignment and applied to the computational model. The potential for crepitation was then measured as a function of tendo-femoral contact during the simulated deep knee bend cycle to 120°. The cumulative tendo-femoral contact area was defined by any contact within 2 mm of the intercondylar notch. Subsequently this model was modified to match the parameters of 3 pairs of patients from the clinical case control study comparing patients with and without patellar crepitus.23) The tendo-femoral contact areas, and the minimum tendon-to-notch distance at 120° of flexion was measured in these modeled patients.
Analysis of the model with variations in the clinical parameters demonstrated an increased cumulative contact area within 2 mm of the intercondylar notch with decreasing patellar tendon length, an extended femoral component posture, an elevated joint line, and use of a smaller patellar component (Fig. 8). Conversely, increasing the patellar tendon length, flexing the femoral component, lowering the joint line and using the largest possible patellar component decreased the cumulative tendo-femoral contact area. Of these parameters, variations in the patellar tendon length had the largest effect on the cumulative tendo-femoral contact area which is supported by the findings of previous clinical studies.4,16,23)
Analysis of the model with patient matched parameters demonstrated significantly different contact areas. Crepitus patient models demonstrated increased tendo-femoral contact near the intercondylar notch, especially along its anterior border, which was not seen in the control group (Fig. 9). Additionally, the tendon-to-notch distances were significantly less in the crepitus patient models in comparison to their matched control counterparts who had not developed patellar crepitus. Lastly, the alterations of the crepitus model patient parameters to test alternate surgical alignments demonstrated that flexing the femoral component and decreasing the joint line dramatically decreased the cumulative contact area near the intercondylar box (Fig. 10).
Notably, analysis of the control matched patient models demonstrated marked differences in contact areas and locations. Crepitus patient models showed increased contact of the distal quadriceps tendon on the anterior edge of the intercondylar box, while control patient models showed none. This gives supporting evidence to the central pathophysiological hypothesis that increased tendo-femoral contact is a contributing factor for the development of fibrosynovial tissue hypertrophy and subsequent patellar crepitus or clunk.
PREVENTATIVE MEASURES
While many authors have demonstrated that painful patellar crepitus or clunk can effectively be treated with either an open or arthroscopic debridement,24-30) prevention is preferred. The authors now counsel patients with previous knee surgery or with a decreased preoperative patellar tendon length that they are at increased risk for the development of patellar crepitus.
Selection of PS femoral components with a lower intercondylar box ratio (trochlear groove extended more posteriorly and distally) has been shown to lessen patellar crepitus and clunk.11) It is wise to avoid use of an undersized femoral component positioned in a flexed posture.4,23) Excessive femoral component flexion can be prevented by placing the entry site for intramedullary distal femoral cutting guide jig more anteriorly. Due to the anterior bow of the femur, a more posterior entry site may lead the surgeon to cut the distal femur in excessive flexion.
As previously stated, a preoperative diagnosis of patella baja may portend postoperative patella baja, in which case risk counseling is appropriate. Surgeons can avoid creating patella baja by avoiding excessive distal femoral resection. If an excessive amount of distal femur is removed, an increased tibial bearing thickness is required to adequately tension the extension gap, which raises the joint line and creates patella baja.
Yau et al.16) demonstrated a higher chance of developing patellar crepitus with anterior placement of the tibial tray although this surgical variance is not supported by the authors' investigation.23) Nevertheless, neutral or posterior placement of the tibial tray may lessen the development numerous patellar complications including crepitus and clunk.
Normal patellae contact the trochlear groove around 20° of flexion, and as knee flexion increases the contract stresses increase and shift superiorly. Therefore, the patellar component should be placed as superior as possible to avoid unresurfaced bone contact with the femoral component, but not extend superiorly beyond the superior border of the patella which can result in quadriceps tendon irritation and the subsequent development of patellar crepitus or clunk. Additionally, creation of a thinner composite thickness and use of a smaller (thinner) patellar component should be avoided which reduces the offset of the quadriceps tendon from the top of the trochlear groove.23,31) Thus, surgeons should be careful to avoid over-resecting the patella and should use the largest possible patellar component that fits the native bone stock. Meftah et al.,3) further demonstrated a reduction in peripatellar crepitation from 40% to 21% by avoiding overstuffing the patellofemoral joint and resecting bone that is not covered by the patellar component. The senior author now routinely resects any uncovered lateral patellar facet to avoid postoperative anterior knee pain and patellar crepitus.
Lastly, the quadriceps tendon often has a proliferation of fibrosynovial tissue at the border of the superior pole of the patella and distal quadriceps tendon at the time of primary TKA. It is this tissue that has been implicated in the pathophysiology of patellar crepitus and clunk. The senior author excises this tissue routinely to avoid the development of patellofemoral crepitus and clunk (Fig. 11).
The excellent clinical results of patients with PS TKA are well documented. However, patellar crepitus and clunk syndrome remain a problem in a minority of patients. The cause of patellar crepitus is multifactorial with a myriad of proposed etiologies. Through acknowledgement of these etiological variables, use of modern femoral components with extended trochlear groove geometry, and avoidance of intraoperative errors, patellar crepitus and patellar clunk syndrome can be effectively minimized.
Footnotes
No potential conflict of interest relevant to this article was reported.
References
- 1.Wing CK, Kwok-Hing C. Sixteen years' result of posterior-stabilized TKA. J Knee Surg. 2012;25(3):245–248. doi: 10.1055/s-0031-1299655. [DOI] [PubMed] [Google Scholar]
- 2.Thadani PJ, Vince KG, Ortaaslan SG, Blackburn DC, Cudiamat CV. Ten- to 12-year followup of the Insall-Burstein I total knee prosthesis. Clin Orthop Relat Res. 2000;(380):17–29. doi: 10.1097/00003086-200011000-00004. [DOI] [PubMed] [Google Scholar]
- 3.Meftah M, Ranawat AS, Ranawat CS. The natural history of anterior knee pain in 2 posterior-stabilized, modular total knee arthroplasty designs. J Arthroplasty. 2011;26(8):1145–1148. doi: 10.1016/j.arth.2010.12.013. [DOI] [PubMed] [Google Scholar]
- 4.Figgie HE, 3rd, Goldberg VM, Heiple KG, Moller HS, 3rd, Gordon NH. The influence of tibial-patellofemoral location on function of the knee in patients with the posterior stabilized condylar knee prosthesis. J Bone Joint Surg Am. 1986;68(7):1035–1040. [PubMed] [Google Scholar]
- 5.Insall JN, Lachiewicz PF, Burstein AH. The posterior stabilized condylar prosthesis: a modification of the total condylar design: two to four-year clinical experience. J Bone Joint Surg Am. 1982;64(9):1317–1323. [PubMed] [Google Scholar]
- 6.Hozack WJ, Rothman RH, Booth RE, Jr, Balderston RA. The patellar clunk syndrome: a complication of posterior stabilized total knee arthroplasty. Clin Orthop Relat Res. 1989;(241):203–208. [PubMed] [Google Scholar]
- 7.Niikura T, Tsumura N, Tsujimoto K, Yoshiya S, Kurosaka M, Shiba R. Patellar clunk syndrome after TKA with cruciate retaining design: a report of two cases. Orthopedics. 2008;31(1):90. doi: 10.3928/01477447-20080101-30. [DOI] [PubMed] [Google Scholar]
- 8.Sringari T, Maheswaran SS. Patellar clunk syndrome in patellofemoral arthroplasty: a case report. Knee. 2005;12(6):456–457. doi: 10.1016/j.knee.2004.11.008. [DOI] [PubMed] [Google Scholar]
- 9.Beight JL, Yao B, Hozack WJ, Hearn SL, Booth RE., Jr The patellar "clunk" syndrome after posterior stabilized total knee arthroplasty. Clin Orthop Relat Res. 1994;(299):139–142. [PubMed] [Google Scholar]
- 10.Clarke HD, Fuchs R, Scuderi GR, Mills EL, Scott WN, Insall JN. The influence of femoral component design in the elimination of patellar clunk in posterior-stabilized total knee arthroplasty. J Arthroplasty. 2006;21(2):167–171. doi: 10.1016/j.arth.2005.05.024. [DOI] [PubMed] [Google Scholar]
- 11.Fukunaga K, Kobayashi A, Minoda Y, Iwaki H, Hashimoto Y, Takaoka K. The incidence of the patellar clunk syndrome in a recently designed mobile-bearing posteriorly stabilised total knee replacement. J Bone Joint Surg Br. 2009;91(4):463–468. doi: 10.1302/0301-620X.91B4.21494. [DOI] [PubMed] [Google Scholar]
- 12.Ip D, Ko PS, Lee OB, Wu WC, Lam JJ. Natural history and pathogenesis of the patella clunk syndrome. Arch Orthop Trauma Surg. 2004;124(9):597–602. doi: 10.1007/s00402-003-0533-9. [DOI] [PubMed] [Google Scholar]
- 13.Lonner JH, Jasko JG, Bezwada HP, Nazarian DG, Booth RE., Jr Incidence of patellar clunk with a modern posterior-stabilized knee design. Am J Orthop (Belle Mead NJ) 2007;36(10):550–553. [PubMed] [Google Scholar]
- 14.Ranawat AS, Ranawat CS, Slamin JE, Dennis DA. Patellar crepitation in the P.F.C. sigma total knee system. Orthopedics. 2006;29(9 Suppl):S68–S70. [PubMed] [Google Scholar]
- 15.Schroer WC, Diesfeld PJ, Reedy ME, LeMarr A. Association of increased knee flexion and patella clunk syndrome after mini-subvastus total knee arthroplasty. J Arthroplasty. 2009;24(2):281–287. doi: 10.1016/j.arth.2007.10.005. [DOI] [PubMed] [Google Scholar]
- 16.Yau WP, Wong JW, Chiu KY, Ng TP, Tang WM. Patellar clunk syndrome after posterior stabilized total knee arthroplasty. J Arthroplasty. 2003;18(8):1023–1028. doi: 10.1016/s0883-5403(03)00447-9. [DOI] [PubMed] [Google Scholar]
- 17.Pollock DC, Ammeen DJ, Engh GA. Synovial entrapment: a complication of posterior stabilized total knee arthroplasty. J Bone Joint Surg Am. 2002;84(12):2174–2178. [PubMed] [Google Scholar]
- 18.Ip D, Wu WC, Tsang WL. Comparison of two total knee prostheses on the incidence of patella clunk syndrome. Int Orthop. 2002;26(1):48–51. doi: 10.1007/s00264-001-0316-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Maloney WJ, Schmidt R, Sculco TP. Femoral component design and patellar clunk syndrome. Clin Orthop Relat Res. 2003;(410):199–202. doi: 10.1097/01.blo.0000063606.67412.96. [DOI] [PubMed] [Google Scholar]
- 20.Anderson MJ, Becker DL, Kieckbusch T. Patellofemoral complications after posterior-stabilized total knee arthroplasty: a comparison of 2 different implant designs. J Arthroplasty. 2002;17(4):422–426. doi: 10.1054/arth.2002.32173. [DOI] [PubMed] [Google Scholar]
- 21.Shoji H, Shimozaki E. Patellar clunk syndrome in total knee arthroplasty without patellar resurfacing. J Arthroplasty. 1996;11(2):198–201. doi: 10.1016/s0883-5403(05)80017-8. [DOI] [PubMed] [Google Scholar]
- 22.Frye BM, Floyd MW, Pham DC, Feldman JJ, Hamlin BR. Effect of femoral component design on patellofemoral crepitance and patella clunk syndrome after posterior-stabilized total knee arthroplasty. J Arthroplasty. 2012;27(6):1166–1170. doi: 10.1016/j.arth.2011.12.009. [DOI] [PubMed] [Google Scholar]
- 23.Dennis DA, Kim RH, Johnson DR, Springer BD, Fehring TK, Sharma A. The John Insall Award: control-matched evaluation of painful patellar Crepitus after total knee arthroplasty. Clin Orthop Relat Res. 2011;469(1):10–17. doi: 10.1007/s11999-010-1485-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24.Dajani KA, Stuart MJ, Dahm DL, Levy BA. Arthroscopic treatment of patellar clunk and synovial hyperplasia after total knee arthroplasty. J Arthroplasty. 2010;25(1):97–103. doi: 10.1016/j.arth.2008.11.005. [DOI] [PubMed] [Google Scholar]
- 25.Koh YG, Kim SJ, Chun YM, Kim YC, Park YS. Arthroscopic treatment of patellofemoral soft tissue impingement after posterior stabilized total knee arthroplasty. Knee. 2008;15(1):36–39. doi: 10.1016/j.knee.2007.08.009. [DOI] [PubMed] [Google Scholar]
- 26.Wong JW, Yau PW, Chiu PK. Arthroscopic treatment of patellar symptoms in posterior stabilized total knee replacement. Int Orthop. 2002;26(4):250–252. doi: 10.1007/s00264-002-0347-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Takahashi M, Miyamoto S, Nagano A. Arthroscopic treatment of soft-tissue impingement under the patella after total knee arthroplasty. Arthroscopy. 2002;18(4):E20. doi: 10.1053/jars.2002.31968. [DOI] [PubMed] [Google Scholar]
- 28.Lucas TS, DeLuca PF, Nazarian DG, Bartolozzi AR, Booth RE., Jr Arthroscopic treatment of patellar clunk. Clin Orthop Relat Res. 1999;(367):226–229. [PubMed] [Google Scholar]
- 29.Messieh M. Management of patellar clunk under local anesthesia. J Arthroplasty. 1996;11(2):202–203. doi: 10.1016/s0883-5403(05)80018-x. [DOI] [PubMed] [Google Scholar]
- 30.Vernace JV, Rothman RH, Booth RE, Jr, Balderston RA. Arthroscopic management of the patellar clunk syndrome following posterior stabilized total knee arthroplasty. J Arthroplasty. 1989;4(2):179–182. doi: 10.1016/s0883-5403(89)80072-5. [DOI] [PubMed] [Google Scholar]
- 31.Hoops HE, Johnson DR, Kim RH, et al. Control-matched computational evaluation of tendo-femoral contact in patients with posterior-stabilized total knee arthroplasty. J Orthop Res. 2012;30(9):1355–1361. doi: 10.1002/jor.22060. [DOI] [PubMed] [Google Scholar]