Skip to main content
. 2014 Feb;6(2):126–133. doi: 10.3978/j.issn.2072-1439.2014.02.01

Table 4. Diagnosis at the optimal cutoff points of the QRS amplitude and Tslew.

Index (%) % (95% CI)
Se Sp PV+ PV– Coincidence rate
QRS ≤72.3 68.42 (43.45, 87.42) 90.91 (58.72, 99.77) 92.86 (66.13, 99.82) 62.50 (35.43, 84.80) 76.67 (57.72, 90.07)
Tslew ≤90 94.74 (73.97, 99.87) 81.82 (48.22,97.72) 90.00 (68.30, 98.77) 90.00 (55.50, 99.75) 90.00 (73.47, 97.89)

QRS, QRS amplitude; Tslew, the maximum slope of the descending T wave of ventricular evoked response (VER); Se, sensitivity; Sp, specificity; PV+, positive predictive values; PV–, negative predictive values; CI, confidence interval.