Table 1.

Microbial contamination of dental unit waterlines.

Control Intermittent Disinfection Continuous Disinfection
Microbiological indicators untreated hydrogen peroxide, peracetic acid 0.26% hydrogen peroxide, peracetic acid 0.26% hydrogen peroxide, silver ions 0.02% stabilized ClO2 0.22%
mains water n: 18 mains water n: 18 deionized water n: 18 deionized water n: 18 reverse osmosis n: 18
HPC 22 °C
positive samples (%) 100% 100% 100% 22.2% 55.6%
range (log10 CFU/mL) (2.00–3.87) (1.30–2.96) (0.30–3.86) (0.00–1.00) (0.00–1.85)
mean ± SD (log10 CFU/mL) 3.11 ± 0.47 1.95 ± 0.60 2.20 ± 1.18 0.14 ± 0.30 0.68 ± 0.73
HPC 36 °C
positive samples (%) 100% 100% 94.14% 27.8% 100%
range (log10 CFU/mL) (3.00–4.00) (0.56–2.96) (0.00–3.57) (0.00–1.04) (0.30–2.20)
mean ± SD (log10 CFU/mL) 3.41 ± 0.37 2.26 ± 0.43 2.30 ± 1.01 0.19 ± 0.36 1.06 ± 0.60
Pseudomonas aeruginosa
positive samples (%) 27.8% 55.6% 27.8% absent in all samples absent in all samples
range (log10 CFU/100 mL) (0.00–1.49) (0.00–1.41) (0.00–3.00)
mean ± SD (log10 CFU/100 mL) 0.17 ± 0.38 0.55 ± 0.60 0.61 ± 1.05