Table 1.
Ref. | Country of Origin | Age (yr)/Gender | Reason for stent placement | Location | Pre-stent procedures | Initial stent | Stent fracture time after placement | Stent fracture management | Repeat stent placement |
Current Case | United States | 71/female | Dysphagia from metastatic esophageal squamous cell carcinoma | Mid esophagus | None | 18 mm × 120 mm, Fully covered Nitinol SEMS - Bonastent, Standard Sci-Tech, Inc, Seoul, South Korea | 45 wk | Proximal piece in the upper esophagus and distal fragment wedged in a hiatal hernia, both removed endoscopically | 18 mm × 100 mm, Partially covered Nitinol SEMS - Wallflex, Boston Scientific, Marlborough, MA, United States |
Wadsworth et al[8], 2010 | United Kingdom | 75/female | Dysphagia from refractory benign esophageal stricture | Distal esophagus | None | 22 mm × 120 mm, Fully covered Nitinol SEMS - EBN stent, Diagmed Healthcare Ltd, Thirsk, United Kingdom | 8 wk | Proximal piece in esophagus removed endoscopically, and distal fragment migrated to the colon, passed rectally | Was under consideration at the time of publication |
Wiedmann et al[9], 2009 | Germany | 69/not reported | Dysphagia from metastastic EAC | Distal esophagus | None | 22 mm × 160 mm, Fully covered Nitinol SEMS - Hanarostent, M.I.Tech Co., Inc, Seoul, South Korea | 20 wk | Proximal piece in esophagus and distal fragment in the gastric antrum, both removed endoscopically | 22 mm × 120 mm, Fully covered Nitinol SEMS - Hanarostent, M.I.Tech Co., Inc, Seol, South Korea |
Chhetri et al[10], 2008 | United Kingdom | 50/male | Dysphagia from EAC | Distal esophagus | Palliative chemotherapy | 18 mm × 110 mm, Fully covered Nitinol SEMS - Choostent, M.I.Tech Co., Inc, Seoul, South Korea | 28 wk | Proximal piece in the esophagus removed endoscopically as it was causing trauma and bleeding, and the distal fragment wedged in the distal two-thirds of the tumor | 18 mm × 120 mm, Partially covered Nitinol SEMS - Ultraflex Boston Scientific, Natick, MA, United States, was placed across the exposed upper end of the tumor and through the prior fractured stent |
Doğan et al[11], 2005 | Turkey | 50/not reported | Dysphagia from metastastic EAC | Distal esophagus | None | 18 mm × 100 mm, Uncovered Nitinol SEMS - Esophacoil, Medtronic InStent Inc., Minneapolis, MN, Unites States | 10 wk | Proximal piece with tumor overgrowth and partial obstruction in the esophagus and distal fragment migrated and left in the stomach | 18 mm, Uncovered Nitinol SEMS - Esophacoil, Medtronic InStent Inc., Minneapolis, MN, United States inserted through the proximal fractured fragment |
Reddy et al[12], 2003 | United Kingdom | 76/male | Dysphagia from EAC at the GE junction | GE junction | Dilation and argon plasma ablation | 18 mm × 100 mm, Uncovered Nitinol SEMS - Esophacoil, Kimal PLC, Uxbridge, United Kindom | 22 wk | Proximal piece in the esophagus with tumor occlusion and the distal fragment in the stomach which subsequently migrated into the right inguinal hernia, removed surgically by enterotomy, followed by herniorrhaphy | Uncovered Nitinol SEMS - Esophacoil, Medtronic InStent Inc., Minneapolis, MN, United States inserted through the proximal fractured fragment |
Reddy et al[12], 2003 | United Kingdom | 76/male (same patient) | Dysphagia from EAC at the GE junction | GE junction | None | Size NR, Uncovered Nitinol SEMS - Esophacoil, Kimal PLC, Uxbridge, United Kingdom | 23 wk | Stent fracture in 2 places with proximal piece in the esophagus, middle piece in the stomach and the distal piece in the small intestine, not removed as patient died due to aspiration pneumonia | None |
Altiparmak et al[13], 2000 | Turkey | 52/male | Dysphagia from EAC | NR | None | Size NR, Uncovered Nitinol SEMS - Wallstent, Schneider Inc., Plymouth, MN, United States | 40 wk | Proximal piece in the esophagus and the distal fragment in the stomach causing a gastrocolic fistula with unsuccessful endoscopic removal requiring surgical gastrotomy and fistula repair | None |
Grimley et al[14], 1999 | United Kingdom | 79/male | Dysphagia from refractory anastomotic stricture after resection of EAC | Proximal esophagus | Dilation of anastomotic stricture | 18 mm × 100 mm, Uncovered Nitinol SEMS - Esophacoil, Medtronic InStent Inc., Minneapolis, MN, Unites States | 8 wk | Proximal piece in the esophagus and the distal 2 cm fragment migrated into the stomach, subsequently passed rectally | Partially covered Nitinol SEMS - Ultraflex Boston Scientific, Galway, Ireland, was placed across the exposed upper end of the tumor and through the prior fractured stent |
SEMS: Self-expanding metal stent; EAC: Esophageal adenocarcinoma; GE: Gastroesophageal.