TABLE 1.

Magnetic Resonance Imaging Sequence Parameters

Imaging Parameters 2D MESE 3D GRE 3D FLASH
Weighting T2 T1rho T1
Plane Sagittal Sagittal Sagittal
Fat sat Yes Yes Yes
Matrix 256 × 128 256 × 128 320 × 320
Number of slices 31 30 80
FOV (mm) 150 150 130
Slice thickness (mm) 1.5 3 1.5
Flip angle (°) 180 25 3.9/23 (Before); 5.5/32.5 (After)
TE (ms) 16.5/33/49.5/66/82.5 2.04 7.24
TR (ms) 4000 175 17
BW (Hz/pixel) 130 260 130
Interpolated in-plane spatial resolution (mm) 0.59 × 0.59 0.59 × 0.59 0.7 × 0.7
Echo train length 5 1 1
NEX 1 1 2
Duration of each 90° pulse (μs) 1536 200 200
Acceleration factor 2 2
Spin-lock frequency (Hz) 300
Time of spin-lock (TSL) [ms] 2/10/20/30
Acquisition time 8 minutes 30 seconds 5 minutes 31 seconds/TSL <15 minutes

BW, band width; FLASH, fast low angle shot; FOV, field of view; GRE, gradient recalled echo; MESE, multiecho spin echo (MESE); NEX, number of excitations; TE, echo time; TR, relaxation time.