TABLE 1.

Comparison of heterotrophic plate counts for standard and silver-coated needleless connectors

Sample parameter Standard NCs Silver-coated NCs P value
Total no. of samples collected 76 75
No. of samples with deviation from collection protocol 0 4
No. of samples for analysis 76 71
No. (%) of positive samples 35 (46.1) 42 (59.2) 0.11a
No. of samples with >1,000 CFU/NC 4 7 0.456b
No. of NC samples containing ≥2 species/no. of NC samples characterized 5/17 14/29 0.592c
a

Determined using the nonparametric Wilcoxon test.

b

Determined using the chi-square test.

c

Determined using the chi-square test.