Skip to main content
. 2014 Mar;52(3):971–973. doi: 10.1128/JCM.02567-13

TABLE 2.

Clinical performance characteristics of the two commercial PCR assays performed on the cobas z 480 analyzer for the detection of M. genitalium

Assay Result No. of samples by in-house PCR result
% sensitivitya (95% CI) % specificitya (95% CI) NPVb (%) PPVb (%)
Positive Negative
TIB MOLBIOL Positive 50 0 92.6 (82.4–97.1) NA 92.6 100
Negative 4 50 NA 100 (92.9–100)
Diagenode Positive 47 0 87 (75.6–93.4) NA 87.7 100
Negative 7 50 NA 100 (92.9–100)
a

The results of the commercial assays were compared to the in-house results using McNemar's test. P values were >0.05. CI, confidence interval; NA, not applicable.

b

NPV, negative predictive value; PPV, positive predictive value.