Skip to main content
. Author manuscript; available in PMC: 2014 Dec 3.
Published in final edited form as: Anal Chem. 2013 Nov 19;85(23):11650–11657. doi: 10.1021/ac4033615

Table 3.

Accuracies of Retention Projections and Linear Retention Indices in Five Different Labs, Each Running Three Unique Methods

Lab Methoda Retention
projection
accuracy (s)b
Linear retention
indexing
accuracy (s)b
A 1 ±1.9 ±230
A 2 ±1.4 ±22
A 3 ±2.0 ±54

B 4 ±0.28 ±11
B 5 ±0.45 ±11
B 6 ±0.59 ±98

C 7 ±1.5 ±19
C 8 ±0.24 ±4.4
C 9 ±3.8 ±80

D 10 ±0.88 ±3.3
D 11 ±0.74 ±3.5
D 12 ±0.66 ±3.9

E 13 ±1.2 ±21
E 14 ±1.4 ±54
E 15 ±1.1 ±47
Average: ±1.2 ±44
a

Methods are described in Supporting Information.

b

Accuracy was calculated from the root-mean-square of the difference between the experimental and predicted retention times for all 12 test compounds.