Skip to main content
. 2013 May 22;40(6):063902. doi: 10.1118/1.4805098

Table 4.

Penetration through protective garments and lead sheets as a percentage of initial intensity.

  60 kVp
70 kVp
80 kVp
100 kVp
120 kVp
Lead (mm)1 P2 S2 P S P S P S P S
0.30 0.99 5.3 3.9 9.4 7.3 11.6 9.9
0.35 0.51 3.5 2.5 6.9 5.1 8.6 7.2
0.40 0.28 2.4 1.6 5.1 3.7 6.5 5.4
0.45 0.15 1.7 1.1 3.9 2.7 5.0 4.1
0.50 0.10 1.2 0.75 3.0 2.0 3.9 3.2
Garment
P
S
P
S
P
S
P
S
P
S
A (lead) 0.17 0.090 0.57 1.3 0.7 3.0 2.0 3.9 3.1
B (lead-free) 0.24 0.17 0.68 1.6 1.1 3.9 2.6 5.8 4.2
C (lead-free) 0.40 0.43 0.88 1.5 1.5 3.9 3.2 6.6 5.1
1

Penetration calculated from polynomial fits listed in Table 3.

2

P = primary, S = scattered.