Skip to main content
. 2014 Jan;110(1):199–212. doi: 10.1016/j.radonc.2013.11.002

Table 3.

Example 3 – HDR 192Ir BT source for breast balloon applicator.

Category Typical level (%) Assumptions
Source strength 2 PSDL traceable calibrations
Treatment planning 3 Reference data with the appropriate bin width
Medium dosimetric corrections 3 Balloon filled with standard level of contrast agent, no consideration or composition of chestwall, lung, or breast
Scatter dosimetric corrections 7 A non-scalar correction for skin dose (and at points in proximity to the surface near the balloon) is needed, and will require an advanced dose calculation formalism to properly account for radiation scatter conditions in the patient. Use of a single prescription point might be not sufficient
Dose delivery including registration of applicator geometry to anatomy 7 Accurate QA concept for commissioning and constancy checks, especially for source positioning and applicator/ source path geometry, appropriate imaging techniques (either small slice thickness, 3D sequences or combination of different slice orientations), applicator characterization
Interfraction/Intrafraction changes between imaging and dose delivery 7 For one treatment plan per applicator insertion and measures to detect major variations for subsequent fractions
Total dosimetric uncertainty (k = 1) 13 For treatment delivered with the same BT source

Estimated value based on expert discussion.